Dosimetric investigation of LDR brachytherapy ¹⁹²Ir wires by Monte Carlo and TPS calculations.

Purpose: The aim of this study was to investigate the dose rate distribution around (192)Ir wires used as radioactive sources in low-dose-rate brachytherapy applications.Materials and Methods: Monte Carlo modeling of a 0.3-mm diameter source and its surrounding water medium was performed for five di...

Descripción completa

Detalles Bibliográficos
Publicado en:Japanese Journal of Radiology Vol. 31; no. 1; pp. 24 - 31
Autores principales: Bozkurt A, Acun H, Kemikler G, Bozkurt, Ahmet, Acun, Hediye, Kemikler, Gonul
Formato: research Journal Article
Publicado: Springer Nature Jan2013
Acceso en línea:Ver este registro en EBSCOhost
fields @attributes:
  recordID: 1
pdfLink:
plink: https://search.ebscohost.com/login.aspx?direct=true&db=ccm&AN=104295275&site=ehost-live
header:
  @attributes:
    shortDbName: ccm
    uiTerm: 104295275
    longDbName: CINAHL Complete
    uiTag: AN
  controlInfo:
    bkinfo:
    dissinfo:
    jinfo:
      jid:
        18671071
        AUCM
      jtl: Japanese Journal of Radiology
      issn: 18671071
      maglogo: N
    pubinfo:
      dt: Jan2013
      vid: 31
      iid: 1
      pid: 237
      pub: Springer Nature
      place: New York, New York
    artinfo:
      ui:
        104295275
        104295275
        NLM23054886
        2011900882
        10.1007/s11604-012-0140-1
        NLM23054886
        104295275
      ppf: 24
      ppct: 7
      formats:
        fmt:
          @attributes:
            type: P
      tig:
        atl: Dosimetric investigation of LDR brachytherapy ¹⁹²Ir wires by Monte Carlo and TPS calculations.
      aug:
        au:
          Bozkurt A
          Acun H
          Kemikler G
          Bozkurt, Ahmet
          Acun, Hediye
          Kemikler, Gonul
        affil: Department of Physics, Faculty of Arts and Sciences, Harran University, Osmanbey Campus, 63300 Sanliurfa, Turkey
      sug:
        subj:
          Brachytherapy Equipment and Supplies
          Radioisotopes Therapeutic Use
          Radiometry Methods
          Systems Analysis
          Computer Simulation
          Human
          Radiation Dosage
          Radiotherapy, Computer-Assisted
          Software
      ab: Purpose: The aim of this study was to investigate the dose rate distribution around (192)Ir wires used as radioactive sources in low-dose-rate brachytherapy applications.Materials and Methods: Monte Carlo modeling of a 0.3-mm diameter source and its surrounding water medium was performed for five different wire lengths (1-5 cm) using the MCNP software package.Results: The computed dose rates per unit of air kerma at distances from 0.1 up to 10 cm away from the source were first verified with literature data sets. Then, the simulation results were compared with the calculations from the XiO CMS commercial treatment planning system.Conclusion: The study results were found to be in concordance with the treatment planning system calculations except for the shorter wires at close distances.
      pubtype: Academic Journal
      doctype:
        research
        Journal Article
      ougenre: Article
    language: English
    refInfo:
    holdings:
      @attributes:
        islocal: N