Comparison of three multichannel transmit/receive radiofrequency coil configurations for anatomic and functional cardiac MRI at 7.0T: implications for clinical imaging.

Objectives: To implement, examine, and compare three multichannel transmit/receive coil configurations for cardiovascular MR (CMR) at 7T.Methods: Three radiofrequency transmit-receive (TX/RX) coils with 4-, 8-, and 16-coil elements were used. Ten healthy volunteers (seven males, age 28 ± 4 years) un...

Descripción completa

Detalles Bibliográficos
Publicado en:European Radiology Vol. 22; no. 10; pp. 2211 - 2221
Autores principales: Winter L, Kellman P, Renz W, Gräßl A, Hezel F, Thalhammer C, von Knobelsdorff-Brenkenhoff F, Tkachenko V, Schulz-Menger J, Niendorf T, Winter, Lukas, Kellman, Peter, Renz, Wolfgang, Gräßl, Andreas, Hezel, Fabian, Thalhammer, Christof, von Knobelsdorff-Brenkenhoff, Florian, Tkachenko, Valeriy, Schulz-Menger, Jeanette, Niendorf, Thoralf
Formato: research Journal Article
Publicado: Springer Nature Oct2012
Acceso en línea:Ver este registro en EBSCOhost
Descripción
Sumario:Objectives: To implement, examine, and compare three multichannel transmit/receive coil configurations for cardiovascular MR (CMR) at 7T.Methods: Three radiofrequency transmit-receive (TX/RX) coils with 4-, 8-, and 16-coil elements were used. Ten healthy volunteers (seven males, age 28 ± 4 years) underwent CMR at 7T. For all three RX/TX coils, 2D CINE FLASH images of the heart were acquired. Cardiac chamber quantification, signal-to-noise ratio (SNR) analysis, parallel imaging performance assessment, and image quality scoring were performed.Results: Mean total examination time was 29 ± 5 min. All images obtained with the 8- and 16-channel coils were diagnostic. No significant difference in ejection fraction (EF) (P > 0.09) or left ventricular mass (LVM) (P > 0.31) was observed between the coils. The 8- and 16-channel arrays yielded a higher mean SNR in the septum versus the 4-channel coil. The lowest geometry factors were found for the 16-channel coil (mean ± SD 2.3 ± 0.5 for R = 4). Image quality was rated significantly higher (P < 0.04) for the 16-channel coil versus the 8- and 4-channel coils.Conclusions: All three coil configurations are suitable for CMR at 7.0T under routine circumstances. A larger number of coil elements enhances image quality and parallel imaging performance but does not impact the accuracy of cardiac chamber quantification.Key Points: • Cardiac chamber quantification using 7.0T magnetic resonance imaging is feasible. • Examination times for cardiac chamber quantification at 7.0T match current clinical practice. • Multichannel transceiver RF technology facilitates improved image quality and parallel imaging performance. • Increasing the number of RF channels does not influence cardiac chamber quantification.