The digital control of anaesthetic gas flow.

The theory and construction of a prototype digital gas flow controller are described. Using eight preset needle valves, it has the ability to deliver any flow from 50 to 12750 ml/minute in steps of 50 ml/minute. Under given conditions, the accuracy of this device is very high and its variation in pe...

Descripción completa

Detalles Bibliográficos
Publicado en:Anaesthesia Vol. 41; no. 4; pp. 413 - 419
Autores principales: Boaden, R W, Hutton, P
Formato: research Journal Article
Publicado: Wiley-Blackwell Apr1986
Acceso en línea:Ver este registro en EBSCOhost
Descripción
Sumario:The theory and construction of a prototype digital gas flow controller are described. Using eight preset needle valves, it has the ability to deliver any flow from 50 to 12750 ml/minute in steps of 50 ml/minute. Under given conditions, the accuracy of this device is very high and its variation in performance with pipeline supply pressures is quantified. The required flow is requested from a BBC 'B' microcomputer which is interfaced with the equipment via a program written in Basic and the 1MHz bus port. The possible uses and potential of a microcomputer-controlled flow regulator in anaesthesia and intensive care are discussed.