Clinical use of a system for the automated recording and analysis of electrically evoked compound action potentials (ECAPs) in cochlear implant patients.

Conclusion: The CustomSoundEP software can record ECAP traces in an automated manner and calculate the ECAP threshold from them. These values are in good agreement with those determined by audiologists. Manual optimization measurements, as needed when using the predecessor version 3 of the software,...

Descripción completa

Detalles Bibliográficos
Publicado en:Acta Oto-Laryngologica Vol. 130; no. 6; pp. 724 - 733
Autores principales: Gärtner L, Lenarz T, Joseph G, Büchner A
Formato: research tables/charts tracings Journal Article
Publicado: Taylor & Francis Ltd Jun2010
Acceso en línea:Ver este registro en EBSCOhost
Descripción
Sumario:Conclusion: The CustomSoundEP software can record ECAP traces in an automated manner and calculate the ECAP threshold from them. These values are in good agreement with those determined by audiologists. Manual optimization measurements, as needed when using the predecessor version 3 of the software, are done by the software itself in the newer system. Objectives: The programming of a cochlear implant speech processor is usually carried out in a subjective manner, influenced by the audiologist and the subject, and it takes considerable time. With increasing numbers of patients there is a growing need for a rapid and reliable technique to measure and analyse neural responses – one which could give suggestions for programming the speech processor and thus shorten the fitting procedure. Methods: In 30 patients, ECAP responses were recorded postoperatively in an automated procedure. ECAP thresholds were determined automatically by the AutoNRTTM (Automatic Neural Response Telemetry) software and visually by seven persons. Results: The time required to determine a TECAP threshold for a single electrode contact was found to be between 0.5 and 1 min.