The Advanced Bionics High Resolution Mode: stimulation rates up to 5000 pps.

Conclusion: The technology of stimulation rates between 1500 and 5000 pulses per second per channel (pps/ch) brings a benefit for all users. However, the fastest possible rate does not necessarily lead to best hearing performance in each individual subject. Individual optimization of stimulation rat...

Full description

Bibliographic Details
Published in:Acta Oto-Laryngologica Vol. 130; no. 1; pp. 114 - 124
Main Authors: Buechner A, Frohne-Büchner C, Gaertner L, Stoever T, Battmer RD, Lenarz T
Format: equations & formulas research tables/charts Journal Article
Published: Taylor & Francis Ltd Jan2010
Online Access:View this record in EBSCOhost
fields @attributes:
  recordID: 1
pdfLink:
plink: https://search.ebscohost.com/login.aspx?direct=true&db=ccm&AN=105295906&site=ehost-live
header:
  @attributes:
    shortDbName: ccm
    uiTerm: 105295906
    longDbName: CINAHL Complete
    uiTag: AN
  controlInfo:
    bkinfo:
    dissinfo:
    jinfo:
      jid:
        00016489
        B6D
      jtl: Acta Oto-Laryngologica
      issn: 00016489
      maglogo: Y
    pubinfo:
      dt: Jan2010
      vid: 130
      iid: 1
      pid: 377
      pub: Taylor & Francis Ltd
      place: Philadelphia, Pennsylvania
    artinfo:
      ui:
        105295906
        2010539759
        10.3109/00016480902971239
        NLM19479460
        105295906
      ppf: 114
      ppct: 10
      formats:
        fmt:
          @attributes:
            type: P
      tig:
        atl: The Advanced Bionics High Resolution Mode: stimulation rates up to 5000 pps.
      aug:
        au:
          Buechner A
          Frohne-Büchner C
          Gaertner L
          Stoever T
          Battmer RD
          Lenarz T
        affil: Medical University of Hannover
      sug:
        subj:
          Bionics
          Cochlear Implant
          Speech Perception
          Germany
          Human
          Paired T-Tests
          Two-Tailed Test
      ab: Conclusion: The technology of stimulation rates between 1500 and 5000 pulses per second per channel (pps/ch) brings a benefit for all users. However, the fastest possible rate does not necessarily lead to best hearing performance in each individual subject. Individual optimization of stimulation rate is recommended. Objectives: The aim of this study was to investigate the effect of high stimulation rates on speech perception ability. Subjects and methods: Thirteen subjects with at least 3 months of experience with conventional Clarion speech coding strategies were recruited into cross-over studies. The first study investigated stimulation rates of 1500, 2000 and 3000 pps/ch, the second study investigated rates of 2500 and 5000 pps/ch. Speech perception data were collected for testing in quiet and in competing noise . Results: On average, speech perception performance increased significantly when changing from conventional speech coding strategies to High Resolution Mode. Subjects who preferred continous interleaved sampling (CIS) in standard mode appeared to obtain better results with rates of 2500 or 3000 pps/ch than with 5000 pps/ch. The stimulation rate that led to best performance provided the highest stimulation current efficiency in the majority of subjects.
      pubtype: Academic Journal
      doctype:
        equations & formulas
        research
        tables/charts
        Journal Article
      ougenre: Article
    language: English
    refInfo:
    holdings:
      @attributes:
        islocal: N