Performance of bone cements: are current preclinical specifications adequate?

Background Current specifications (standards) for preclinical testing of bone cements (ISO 5833: 2002, ASTM F451-99a) require simple mechanical testing after ageing for 24 h under dry conditions at 23°C. Some bone cements have fulfilled the requirements in the specifications, and yet had inferior cl...

Full description

Bibliographic Details
Published in:Acta Orthopaedica Vol. 79; no. 6; pp. 826 - 832
Main Authors: Nottrott M, Mølster AO, Moldestad IO, Walsh WR, Gjerdet NR
Format: research tables/charts Journal Article
Published: Medical Journals Sweden AB Dec2008
Online Access:View this record in EBSCOhost
fields @attributes:
  recordID: 1
pdfLink:
plink: https://search.ebscohost.com/login.aspx?direct=true&db=ccm&AN=105598948&site=ehost-live
header:
  @attributes:
    shortDbName: ccm
    uiTerm: 105598948
    longDbName: CINAHL Complete
    uiTag: AN
  controlInfo:
    bkinfo:
    dissinfo:
    jinfo:
      jid:
        17453674
        1BVL
      jtl: Acta Orthopaedica
      issn: 17453674
      maglogo: N
    pubinfo:
      dt: Dec2008
      vid: 79
      iid: 6
      pid: 59195
      pub: Medical Journals Sweden AB
    artinfo:
      ui:
        105598948
        2010136439
        10.1080/17453670810016920
        NLM19085502
        105598948
      ppf: 826
      ppct: 6
      formats:
        fmt:
          @attributes:
            type: P
      tig:
        atl: Performance of bone cements: are current preclinical specifications adequate?
      aug:
        au:
          Nottrott M
          Mølster AO
          Moldestad IO
          Walsh WR
          Gjerdet NR
        affil: Department of Orthopaedic Surgery, Haukeland University Hospital,,Section for Orthopaedic Surgery, Department of Surgical Sciences, University of Bergen, Bergen, Norway
      sug:
        subj:
          Bone Cements Standards
          Materials Testing Standards
          Analysis of Variance
          Biophysical Instruments
          Comparative Studies
          Compressive Strength
          Data Analysis Software
          Descriptive Statistics
          Funding Source
          Stress, Mechanical
          Temperature
          Human
      ab: Background Current specifications (standards) for preclinical testing of bone cements (ISO 5833: 2002, ASTM F451-99a) require simple mechanical testing after ageing for 24 h under dry conditions at 23°C. Some bone cements have fulfilled the requirements in the specifications, and yet had inferior clinical results. Clinically, bone cements are subjected to complex loading patterns in a moist or wet environment at 37°C. Thus, both the validity and the robustness of current standard testing protocols can be questioned. Methods We examined the influence of temperature and storage medium on the properties of bone cement. We also compared the results of storage and testing under standard conditions of 23°C in dry air, with the results obtained at 37°C in water or plasma. Results The dry specimens showed an increase in strength and elastic modulus with time, while the values of the wet ones decreased. There was no difference between specimens stored in water or in plasma. Ultimate compressive strength of dry specimens after 24 h was 1.16 times higher than that of the ones stored wet, increasing to 1.34 times after 1 month, and 1.46 times after 6 months (p<0.001 for all comparisons). Interpretation Testing under dry conditions-as required in current standards-always gave higher values for mechanical properties than did storage and testing under more physiological conditions. The sensitivity of test values to different environments implies that testing conditions for bone cements should be scrutinized in order to develop more relevant testing protocols that reflect the in vivo environment more closely.
      pubtype: Academic Journal
      doctype:
        research
        tables/charts
        Journal Article
      ougenre: Article
    language: English
    refInfo:
    holdings:
      @attributes:
        islocal: N