The use of qt interval to determine pacing rate: early clinical experience.

Despite the current availability of physiologic pacing systems, there remain distinct disadvantages associated with their operation. Although in theory the atrial synchronous pacemaker (VDD or DDD) is the ideal system, in practice it performs satisfactorily over a very limited range of heart rates i...

Descripción completa

Detalles Bibliográficos
Publicado en:Pacing & Clinical Electrophysiology Vol. 6; no. 2; pp. 346 - 355
Autores principales: Rickards AF, Donaldson RM, Thalen HJ
Formato: Journal Article
Publicado: Wiley-Blackwell Mar/Apr1983
Acceso en línea:Ver este registro en EBSCOhost
Descripción
Sumario:Despite the current availability of physiologic pacing systems, there remain distinct disadvantages associated with their operation. Although in theory the atrial synchronous pacemaker (VDD or DDD) is the ideal system, in practice it performs satisfactorily over a very limited range of heart rates in a limited number of individuals. We describe the TX system which incorporates ventricular pacing and is driven by T wave sensing determined by the QT interval. This pacemaker can distinguish between physiologic and pathologic tachycardias while providing an almost unlimited rate response even in the presence of an unresponsive atrium.