Changes in the amplitude of endocardial electrograms following defibrillator discharge: comparison of two lead systems.

Changes in the amplitude of endocardial electrograms after an unsuccessful shock attempt have been demonstrated to cause failure of redetection of ventricular fibrillation in patients using an integrated sense-pace defibrillating lead system. Thus, the objective of this study was to compare the effe...

Descripción completa

Detalles Bibliográficos
Publicado en:Pacing & Clinical Electrophysiology Vol. 18; no. 12; pp. 2163 - 2173
Autores principales: Jung W, Manz M, Moosdorf R, Tebbenjohanns J, Pfeiffer D, Lüderitz B
Formato: Journal Article
Publicado: Wiley-Blackwell Dec1995
Acceso en línea:Ver este registro en EBSCOhost
fields @attributes:
  recordID: 1
pdfLink:
plink: https://search.ebscohost.com/login.aspx?direct=true&db=ccm&AN=106038274&site=ehost-live
header:
  @attributes:
    shortDbName: ccm
    uiTerm: 106038274
    longDbName: CINAHL Complete
    uiTag: AN
  controlInfo:
    bkinfo:
    dissinfo:
    jinfo:
      jid:
        01478389
        4F8
      jtl: Pacing & Clinical Electrophysiology
      issn: 01478389
      maglogo: Y
    pubinfo:
      dt: Dec1995
      vid: 18
      iid: 12
      pid: 480
      pub: Wiley-Blackwell
      place: Malden, Massachusetts
    artinfo:
      ui:
        106038274
        2009414179
        NLM8771129
        106038274
      ppf: 2163
      ppct: 10
      formats:
        fmt:
          @attributes:
            type: P
      tig:
        atl: Changes in the amplitude of endocardial electrograms following defibrillator discharge: comparison of two lead systems.
      aug:
        au:
          Jung W
          Manz M
          Moosdorf R
          Tebbenjohanns J
          Pfeiffer D
          Lüderitz B
      sug:
      ab: Changes in the amplitude of endocardial electrograms after an unsuccessful shock attempt have been demonstrated to cause failure of redetection of ventricular fibrillation in patients using an integrated sense-pace defibrillating lead system. Thus, the objective of this study was to compare the effects of defibrillator shocks on the amplitude of endocardial electrograms in 26 patients using two different nonthoracotomy systems, a previous lead (model 0062) or a redesigned version (model 0072). At implant, bipolar endocardial electrograms were obtained before each shock application, during initial detection and redetection of ventricular fibrillation in case the applied shock was unsuccessful, and during intervals of 5, 10, 20, 30, 60, and 120 seconds after each shock delivery. No significant difference was noted in endocardial amplitudes between the lead models 0062 and 0072 during baseline sinus rhythm (12.2 ± 4.6 mV vs 11.4 ± 3.8 mV), and during initial ventricular fibrillation (7.0 ± 2.4 mV vs 7.6 ± 2.3 mV). During redetection of ventricular fibrillation, however, there was a significant difference (P = 0.0006) in endocardial amplitudes (3.4 ± 1.9 mV vs 6.6 ± 2.3 mV) between both leads tested. Comparing lead models 0062 and 0072, marked differences were found in endocardial amplitudes during sinus rhythm 5, 10, and 20 seconds after successful arrhythmia termination: 2.8 ± 1.9 mV vs 8.6 ± 2.9 mV(P < 0.0001), 4.6 ± 2.9 mV vs 9.2 ± 3.2 mV(P = 0.0007), and 6.4 ± 4.0 mV vs 10.5 ± 3.6 mV (P = 0.01). At predischarge testing, failure of redetection of ventricular fibrillation was documented in two patients with the lead model 0062 requiring external defibrillation to restore sinus rhythm. These findings demonstrate a significant less postshock attenuation of the endocardial electrogram amplitudes during persistent ventricular fibrillation after an unsuccessful shock attempt as well as during sinus rhythm immediately following an effective shock delivery using the redesigned lead system model 0072 compared to the electrogram amplitudes obtained in patients using the previous lead model 0062.
      pubtype: Academic Journal
      doctype: Journal Article
      ougenre: Article
    language: English
    refInfo:
    holdings:
      @attributes:
        islocal: N