Young age and wedge stem design are risk factors for periprosthetic fracture after arthroplasty due to hip fracture. A case-control study.

BACKGROUND: It has been suggested that a hip fracture as a primary diagnosis for arthroplasty is a risk factor for periprosthetic fracture. PATIENTS AND METHODS: We compared 16 patients with late periprosthetic fracture of the femur to 48 controls. The primary diagnosis for all patients was hip frac...

Descripción completa

Detalles Bibliográficos
Publicado en:Acta Orthopaedica Vol. 76; no. 1; pp. 56 - 61
Autores principales: Sarvilinna R, Huhtala H, Pajamäki J
Formato: research tables/charts Journal Article
Publicado: Medical Journals Sweden AB Feb2005
Acceso en línea:Ver este registro en EBSCOhost
Descripción
Sumario:BACKGROUND: It has been suggested that a hip fracture as a primary diagnosis for arthroplasty is a risk factor for periprosthetic fracture. PATIENTS AND METHODS: We compared 16 patients with late periprosthetic fracture of the femur to 48 controls. The primary diagnosis for all patients was hip fracture. The date of operation for hip fracture was used to select the control group of patients, all of whom were operated at the same time and in the same hospital district. No other selection criteria were used. Complications at the time of index operation, type of prosthesis, age, sex, BMI, patient's diseases, medication, and surgeon's experience were compared using conditional logistic regression. RESULTS: We found that low age at the time of the hip fracture operation increased the risk of periprosthetic fracture (OR = 4.9, CI = 1.2-20). The polished wedge type of prosthesis was associated with a higher risk of periprosthetic fracture than were other designs (OR = 11, CI = 1.2-97). INTERPRETATION: These findings of risk factors for periprosthetic fracture are new and should be confirmed by further studies using larger numbers of patients.