Optimization of atrial defibrillation with a dual-coil, active pectoral lead system.

INTRODUCTION: Atrial defibrillation can be achieved with standard implantable cardioverter defibrillator (ICD) leads, but the optimal shocking configuration is unknown. The objective of this prospective study was to compare atrial defibrillation thresholds (DFTs) with three shocking configurations t...

Full description

Bibliographic Details
Published in:Journal of Cardiovascular Electrophysiology Vol. 15; no. 7; pp. 790 - 795
Main Authors: Rashba EJ, Shorofsky SR, Peters RW, Gold MR
Format: clinical trial research tables/charts Journal Article
Published: Wiley-Blackwell Jul2004
Online Access:View this record in EBSCOhost
fields @attributes:
  recordID: 1
pdfLink:
plink: https://search.ebscohost.com/login.aspx?direct=true&db=ccm&AN=106648359&site=ehost-live
header:
  @attributes:
    shortDbName: ccm
    uiTerm: 106648359
    longDbName: CINAHL Complete
    uiTag: AN
  controlInfo:
    bkinfo:
    dissinfo:
    jinfo:
      jid:
        10453873
        GSB
      jtl: Journal of Cardiovascular Electrophysiology
      issn: 10453873
      maglogo: Y
    pubinfo:
      dt: Jul2004
      vid: 15
      iid: 7
      pid: 480
      pub: Wiley-Blackwell
      place: Malden, Massachusetts
    artinfo:
      ui:
        106648359
        2005096217
        10.1046/j.1540-8167.2004.03684.x
        NLM15250864
        106648359
      ppf: 790
      ppct: 5
      formats:
        fmt:
          @attributes:
            type: P
      tig:
        atl: Optimization of atrial defibrillation with a dual-coil, active pectoral lead system.
      aug:
        au:
          Rashba EJ
          Shorofsky SR
          Peters RW
          Gold MR
        affil: Division of Cardiology, University of Maryland School of Medicine, 22 South Greene Street, Room N3W77, Baltimore, MD 21201-1595; erashba@medicine.umaryland.edu
      sug:
        subj:
          Atrial Fibrillation Therapy
          Defibrillation Methods
          Defibrillators, Implantable
          Electrodes, Implanted
          Clinical Trials
          Data Analysis, Statistical
          Descriptive Statistics
          Maryland
          Prospective Studies
          Human
      ab: INTRODUCTION: Atrial defibrillation can be achieved with standard implantable cardioverter defibrillator (ICD) leads, but the optimal shocking configuration is unknown. The objective of this prospective study was to compare atrial defibrillation thresholds (DFTs) with three shocking configurations that are available with standard ICD leads. METHODS AND RESULTS: This study was a prospective, randomized, paired comparison of shocking configurations on atrial DFTs in 58 patients. The lead system evaluated was a transvenous defibrillation lead with coils in the superior vena cava (SVC) and right ventricular apex (RV) and a left pectoral pulse generator emulator (Can). In the first 33 patients, atrial DFT was measured with the ventricular triad (RV --> SVC + Can) and unipolar (RV --> Can) shocking pathways. In the next 25 patients, atrial DFT was measured with the ventricular triad and the proximal triad (SVC --> RV + Can) configurations. Delivered energy at DFT was significantly lower with the ventricular triad compared to the unipolar configuration (4.7 +/- 3.7 J vs 10.1 +/- 9.5 J, P < 0.001). Peak voltage and shock impedance also were significantly reduced (P < 0.001). There was no significant difference in DFT energy when the ventricular triad and proximal triad shocking configurations were compared (3.6 +/- 3.0 J vs 3.4 +/- 2.9 J for ventricular and proximal triad, respectively, P = NS). Although shock impedance was reduced by 13% with the proximal triad (P < 0.001), this effect was offset by an increased current requirement (10%). CONCLUSION: The ventricular triad is equivalent or superior to other possible shocking pathways for atrial defibrillation afforded by a dual-coil, active pectoral lead system. Because the ventricular triad is also the most efficacious shocking pathway for ventricular defibrillation, this pathway should be preferred for combined atrial and ventricular defibrillators.
      pubtype: Academic Journal
      doctype:
        clinical trial
        research
        tables/charts
        Journal Article
      ougenre: Article
    language: English
    refInfo:
    holdings:
      @attributes:
        islocal: N