Optimization of cardiac resynchronization therapy by continuous hemodynamic monitoring: a case report.

We report the case of a 55-year-old male patient with severe congestive heart failure in whom both a hemodynamic monitor and a biventricular pacemaker were implanted. The potential usefulness of the monitor to optimize AV delay and heart rate was prospectively evaluated. The AV delay associated with...

Full description

Bibliographic Details
Published in:Journal of Cardiovascular Electrophysiology Vol. 15; no. 1; pp. 94 - 97
Main Authors: Braunschweig F, Kjellström B, Gadler F, Linde C
Format: research tables/charts Journal Article
Published: Wiley-Blackwell Jan2004
Online Access:View this record in EBSCOhost
fields @attributes:
  recordID: 1
pdfLink:
plink: https://search.ebscohost.com/login.aspx?direct=true&db=ccm&AN=106673455&site=ehost-live
header:
  @attributes:
    shortDbName: ccm
    uiTerm: 106673455
    longDbName: CINAHL Complete
    uiTag: AN
  controlInfo:
    bkinfo:
    dissinfo:
    jinfo:
      jid:
        10453873
        GSB
      jtl: Journal of Cardiovascular Electrophysiology
      issn: 10453873
      maglogo: Y
    pubinfo:
      dt: Jan2004
      vid: 15
      iid: 1
      pid: 480
      pub: Wiley-Blackwell
      place: Malden, Massachusetts
    artinfo:
      ui:
        106673455
        2004204069
        10.1046/j.1540-8167.2004.03208.x
        NLM15028081
        106673455
      ppf: 94
      ppct: 3
      formats:
        fmt:
          @attributes:
            type: P
      tig:
        atl: Optimization of cardiac resynchronization therapy by continuous hemodynamic monitoring: a case report.
      aug:
        au:
          Braunschweig F
          Kjellström B
          Gadler F
          Linde C
        affil: Department of Cardiology, Karolinska Hospital, 171 76 Stockholm, Sweden; frieder.braunschweig@ks.se
      sug:
        subj:
          Cardiac Pacing, Artificial Methods
          Heart Failure Therapy
          Hemodynamics
          Monitoring, Physiologic Utilization
          Heart Rate
          Male
          Middle Age
          Human
          Middle Aged: 45-64 years
          Male
      ab: We report the case of a 55-year-old male patient with severe congestive heart failure in whom both a hemodynamic monitor and a biventricular pacemaker were implanted. The potential usefulness of the monitor to optimize AV delay and heart rate was prospectively evaluated. The AV delay associated with the lowest filling pressures recorded by the monitor was similar to the optimal AV delay as determined by standard echocardiography. Heart rates >70 beats/min acutely improved hemodynamics but were not tolerated when programmed chronically. This case illustrates that continuous hemodynamic monitoring may offer a useful tool for device tuning and treatment monitoring in cardiac resynchronization therapy.
      pubtype: Academic Journal
      doctype:
        research
        tables/charts
        Journal Article
      ougenre: Article
    language: English
    refInfo:
    holdings:
      @attributes:
        islocal: N