New integrated sensor pacemaker: comparison of rate responses between an integrated minute ventilation and activity sensor and single sensor modes during exercise and daily activities and nonphysiological inference.

A dual sensor DDDR pacemaker (DX2 Model 7970, Medtronic Inc.) has integrated the rate response of minute ventilation (MV) and activity (ACT) sensors. False rate acceleration by the ACT (constrained upper rate) and MV (cross-checked by ACT) is reduced. We examined the rate response profile and rate k...

Descripción completa

Detalles Bibliográficos
Publicado en:Pacing & Clinical Electrophysiology Vol. 19; no. 11; pp. 1664 - 1672
Autores principales: Leung S, Lau C, Tang M, Leung Z
Formato: proceedings research tables/charts Journal Article
Publicado: Wiley-Blackwell Nov1996
Acceso en línea:Ver este registro en EBSCOhost
fields @attributes:
  recordID: 1
pdfLink:
plink: https://search.ebscohost.com/login.aspx?direct=true&db=ccm&AN=107316137&site=ehost-live
header:
  @attributes:
    shortDbName: ccm
    uiTerm: 107316137
    longDbName: CINAHL Complete
    uiTag: AN
  controlInfo:
    bkinfo:
    dissinfo:
    jinfo:
      jid:
        01478389
        4F8
      jtl: Pacing & Clinical Electrophysiology
      issn: 01478389
      maglogo: Y
    pubinfo:
      dt: Nov1996
      vid: 19
      iid: 11
      pid: 480
      pub: Wiley-Blackwell
      place: Malden, Massachusetts
    artinfo:
      ui:
        107316137
        107316137
        1997011406
        10.1111/j.1540-8159.1996.tb03203.x
        NLM8945021
        107316137
      ppf: 1664
      ppct: 8
      formats:
        fmt:
          @attributes:
            type: P
      tig:
        atl: New integrated sensor pacemaker: comparison of rate responses between an integrated minute ventilation and activity sensor and single sensor modes during exercise and daily activities and nonphysiological inference.
      aug:
        au:
          Leung S
          Lau C
          Tang M
          Leung Z
      sug:
        subj:
          Cardiac Pacing, Artificial Methods
          Activities of Daily Living
          Exercise
          Cardiac Pacing, Artificial Equipment and Supplies
          Pacemaker, Artificial
          Exercise Test, Cardiopulmonary
          Equipment Design
          T-Tests
          Descriptive Statistics
          Aged
          Female
          Male
          Human
          Aged: 65+ years
          Female
          Male
      ab: A dual sensor DDDR pacemaker (DX2 Model 7970, Medtronic Inc.) has integrated the rate response of minute ventilation (MV) and activity (ACT) sensors. False rate acceleration by the ACT (constrained upper rate) and MV (cross-checked by ACT) is reduced. We examined the rate response profile and rate kinetics of the automatically optimized integrated sensor by comparing with the projected rate response of ACT and MV sensors alone in nine patients. After 1 month of sensor optimization using rate profile optimization (RPO), patients underwent maximal and submaximal treadmill exercises and performed activities of daily living (ADL). The integrated sensor mode gave a faster speed of rate response with a shorter delay time, time to 50% rate response and time to 90% of rate response compared to the MV sensor during hall walk (0.37 +/- 0.08, 0.7 +/- 0.09, 1.43 +/- 0.19 vs 1.11 +/- 0.1, 1.75 +/- 0.14, 2.91 +/- 0.17 min; P < 0.05). The average maximal sensor rates were significantly more proportional for the integrated sensor mode compared with either the ACT or MV mode. There was no significant difference in both the maximal pacing rate among the three sensor modes during maximal exercise and the rate decay during recovery. During interference studies by arm swinging (30-40 swings/min) and external tapping of the pacemakers (2 taps/s), there was only a moderate increase in pacing rate by 13 +/- 9, 16 +/- 5 beats/min. Hence, the new integrated sensor with the automatic rate profile optimization algorithm resulted in improved rate response profiles during submaximal exercise and ADL compared to the individual sensor response, and the sensor blending and cross-checking algorithm made the pacemaker relatively immune to false triggering of both the ACT and MV sensors.
      pubtype: Academic Journal
      doctype:
        proceedings
        research
        tables/charts
        Journal Article
      ougenre: Article
    language: English
    refInfo:
    holdings:
      @attributes:
        islocal: N