Automated performance evaluation of real-time QRS-detection devices.

This paper presents a QRS-detector error logger and the implementation of a fully automated system for evaluation of QRS-detector performance. The error logger requires a known reference, such as an electrocardiogram (ECG) database, that provides ECG data and reference pulses wherever QRS complexes...

Descripción completa

Detalles Bibliográficos
Publicado en:Biomedical Instrumentation & Technology Vol. 29; no. 1; pp. 41 - 50
Autores principales: Suppappola S, Sun Y, Suppappola, S, Sun, Y
Formato: algorithm research tables/charts Journal Article
Publicado: Association for the Advancement of Medical Instrumentation (AAMI) 1995 Jan-Feb
Acceso en línea:Ver este registro en EBSCOhost
fields @attributes:
  recordID: 1
pdfLink:
plink: https://search.ebscohost.com/login.aspx?direct=true&db=ccm&AN=107415158&site=ehost-live
header:
  @attributes:
    shortDbName: ccm
    uiTerm: 107415158
    longDbName: CINAHL Complete
    uiTag: AN
  controlInfo:
    bkinfo:
    dissinfo:
    jinfo:
      jid:
        08998205
        FZO
      jtl: Biomedical Instrumentation & Technology
      issn: 08998205
      maglogo: N
    pubinfo:
      dt: 1995 Jan-Feb
      vid: 29
      iid: 1
      pid: 13730
      pub: Association for the Advancement of Medical Instrumentation (AAMI)
      place: Arlington, Virginia
    artinfo:
      ui:
        107415158
        107415158
        NLM7704081
        1995020332
        NLM7704081
        107415158
      ppf: 41
      ppct: 9
      formats:
      tig:
        atl: Automated performance evaluation of real-time QRS-detection devices.
      aug:
        au:
          Suppappola S
          Sun Y
          Suppappola, S
          Sun, Y
        affil: University of Rhode Island, Department of Electrical Engineering, Kingston 02881-0805, USA
      sug:
        subj:
          Algorithms
          Diagnosis, Computer Assisted
          Electrocardiography
          QRS Complex
          Software
          Software Design
          Biophysical Instruments
          Human
      ab: This paper presents a QRS-detector error logger and the implementation of a fully automated system for evaluation of QRS-detector performance. The error logger requires a known reference, such as an electrocardiogram (ECG) database, that provides ECG data and reference pulses wherever QRS complexes occur. These reference pulses are used to validate detection pulses generated by a QRS detector in real time. Using an eight-bit microcontroller chip, the error logger was constructed at a very low materials cost. Applications include automated analysis of QRS-detector performance and optimization of parameters in QRS-detection algorithms.