Effect of Applying Force to Self-Adhesive Electrodes on Transthoracic Impedance: Implications for Electrical Cardioversion.

Background Current guidelines disagree on the role for applying force to electrodes during electrical cardioversion (ECV) for atrial fibrillation, particularly when using self-adhesive pads. We evaluated the impact of this practice on transthoracic impedance (TTI) with varying force and in individua...

Full description

Bibliographic Details
Published in:Pacing & Clinical Electrophysiology Vol. 39; no. 10; pp. 1141 - 1148
Main Authors: RAMIREZ, F. DANIEL, FISET, SANDRA L., CLELAND, MARK J., ZAKUTNEY, TIMOTHY J., NERY, PABLO B., NAIR, GIRISH M., REDPATH, CALUM J., SADEK, MOUHANNAD M., BIRNIE, DAVID H.
Format: research tables/charts Journal Article
Published: Wiley-Blackwell Oct2016
Online Access:View this record in EBSCOhost
fields @attributes:
  recordID: 1
pdfLink:
plink: https://search.ebscohost.com/login.aspx?direct=true&db=ccm&AN=118669178&site=ehost-live
header:
  @attributes:
    shortDbName: ccm
    uiTerm: 118669178
    longDbName: CINAHL Complete
    uiTag: AN
  controlInfo:
    bkinfo:
    dissinfo:
    jinfo:
      jid:
        01478389
        4F8
      jtl: Pacing & Clinical Electrophysiology
      issn: 01478389
      maglogo: Y
    pubinfo:
      dt: Oct2016
      vid: 39
      iid: 10
      pid: 480
      pub: Wiley-Blackwell
      place: Malden, Massachusetts
    artinfo:
      ui:
        118669178
        118669178
        118669178
        10.1111/pace.12937
        118669178
      ppf: 1141
      ppct: 7
      formats:
        fmt:
          – @attributes:
              type: T
          – @attributes:
              type: P
      tig:
        atl: Effect of Applying Force to Self-Adhesive Electrodes on Transthoracic Impedance: Implications for Electrical Cardioversion.
      aug:
        au:
          RAMIREZ, F. DANIEL
          FISET, SANDRA L.
          CLELAND, MARK J.
          ZAKUTNEY, TIMOTHY J.
          NERY, PABLO B.
          NAIR, GIRISH M.
          REDPATH, CALUM J.
          SADEK, MOUHANNAD M.
          BIRNIE, DAVID H.
        affil: Arrhythmia Service, University of Ottawa Heart Institute, Ottawa Ontario, Canada
      sug:
        subj:
          Cardiography, Impedance
          Cardioversion Methods
          Atrial Fibrillation Therapy
          Body Mass Index
          Electrodes
          Descriptive Statistics
          Data Analysis Software
          Wilcoxon Rank Sum Test
          P-Value
          Pearson's Correlation Coefficient
          Spearman's Rank Correlation Coefficient
          Repeated Measures
          Analysis of Variance
          Post Hoc Analysis
          Human
          Female
          Male
          Adult
          Middle Age
          Correlational Studies
          Research Protocols
          Inpatients
          Academic Medical Centers Canada
          Canada
          Funding Source
          Adult: 19-44 years
          Middle Aged: 45-64 years
          Female
          Male
      ab: Background Current guidelines disagree on the role for applying force to electrodes during electrical cardioversion (ECV) for atrial fibrillation, particularly when using self-adhesive pads. We evaluated the impact of this practice on transthoracic impedance (TTI) with varying force and in individuals with differing body mass indices (BMI). We additionally assessed whether specific prompts could improve physicians' ECV technique. Methods The study comprised three parts: (1) TTI was measured in 11 participants throughout the respiratory cycle and with variable force applied to self-adhesive electrodes in anteroposterior (AP) and anterolateral (AL) configurations. (2) Three participants in different BMI classes then had TTI measured with prespecified incremental force applied. (3) Ten blinded cardiology trainees simulated ECV on one participant with and without prompting (guideline reminders and force analogies) while force applied and TTI were measured. Results The AP approach was associated with 13% lower TTI than AL (P < 0.001). Strongly negative correlations were observed between force applied and TTI in the AL position, irrespective of BMI (P ≤ 0.003). In all cases, 80% of the total reduction in TTI observed was achieved with 8 kg-force (∼80 N). All prompts resulted in significantly greater force applied and modest reductions in TTI. Conclusions Applying force to self-adhesive electrodes reduces TTI and should be considered as a means of improving ECV success. Numerically greater mean force applied with a 'push-up' force analogy suggests that 'concrete' cues may be useful in improving ECV technique.
      pubtype: Academic Journal
      doctype:
        research
        tables/charts
        Journal Article
      ougenre: Article
    language: English
    refInfo:
    holdings:
      @attributes:
        islocal: N