Effect of using a Planecta™ port with a three-way stopcock on the natural frequency of blood pressure transducer kits.

Blood pressure transducer kits are equipped with two types of Planecta™ ports-the flat-type Planecta™ port (FTP) and the Planecta™ port with a three-way stopcock (PTS). We reported that FTP application decreased the natural frequency of the kits. However, Planecta™ is an invaluable tool as it preven...

Descripción completa

Detalles Bibliográficos
Publicado en:Journal of Clinical Monitoring & Computing Vol. 30; no. 6; pp. 925 - 932
Autores principales: Fujiwara, Shigeki, Tachihara, Keiichi, Mori, Satoshi, Ouchi, Kentaro, Yokoe, Chizuko, Imaizumi, Uno, Morimoto, Yoshinari, Miki, Yoichiro, Toyoguchi, Izumi, Yoshida, Kazu-ichi, Yokoyama, Takeshi
Formato: Journal Article
Publicado: Springer Nature Dec2016
Acceso en línea:Ver este registro en EBSCOhost
fields @attributes:
  recordID: 1
pdfLink:
plink: https://search.ebscohost.com/login.aspx?direct=true&db=ccm&AN=120548705&site=ehost-live
header:
  @attributes:
    shortDbName: ccm
    uiTerm: 120548705
    longDbName: CINAHL Complete
    uiTag: AN
  controlInfo:
    bkinfo:
    dissinfo:
    jinfo:
      jid:
        13871307
        OHC
      jtl: Journal of Clinical Monitoring & Computing
      issn: 13871307
      maglogo: N
    pubinfo:
      dt: Dec2016
      vid: 30
      iid: 6
      pid: 237
      pub: Springer Nature
      place: New York, New York
    artinfo:
      ui:
        120548705
        120548705
        NLM26467334
        10.1007/s10877-015-9795-3
        NLM26467334
        120548705
      ppf: 925
      ppct: 7
      formats:
        fmt:
          @attributes:
            type: P
      tig:
        atl: Effect of using a Planecta™ port with a three-way stopcock on the natural frequency of blood pressure transducer kits.
      aug:
        au:
          Fujiwara, Shigeki
          Tachihara, Keiichi
          Mori, Satoshi
          Ouchi, Kentaro
          Yokoe, Chizuko
          Imaizumi, Uno
          Morimoto, Yoshinari
          Miki, Yoichiro
          Toyoguchi, Izumi
          Yoshida, Kazu-ichi
          Yokoyama, Takeshi
        affil: Department of Anesthesiology, Graduate School of Dentistry , Kanagawa Dental University , 82 Inaoka-cho Yokosuka-shi 238-0003 Japan
      sug:
        subj:
          Blood Pressure Devices
          Transducers
          Blood Pressure Determination Methods
          Catheterization, Peripheral Equipment and Supplies
          Blood Pressure Determination Equipment and Supplies
          Catheterization, Peripheral Methods
          Equipment Design
          Japan
      ab: Blood pressure transducer kits are equipped with two types of Planecta™ ports-the flat-type Planecta™ port (FTP) and the Planecta™ port with a three-way stopcock (PTS). We reported that FTP application decreased the natural frequency of the kits. However, Planecta™ is an invaluable tool as it prevents infection, ensures technical simplicity, and excludes air. Hence, an ideal Planecta™ port that does not decrease the frequency characteristics is required. As a first step in this direction, we aimed to assess the influence of PTSs on the natural frequency of blood transducer kits. A DTXplus transducer kit (DT4812J; Argon Medical Devices, TX, USA) was used along with ≥1 PTSs (JMS, Hiroshima, Japan), and the frequency characteristics were assessed. The natural frequency and damping coefficient of each kit were obtained by using frequency characteristics analysis software, and these parameters were evaluated by plotting them on Gardner's chart. Regardless of whether one or two PTSs were inserted, the natural frequency of the kits only slightly decreased (from 42.5 to 41.1 Hz, when 2 PTSs were used). Thus, the frequency characteristics of the kits with PTSs were adequate for pressure monitoring. The insertion of ≥2 FTPs in pressure transducer kits should be avoided, as they markedly decrease the natural frequency and lead to underdamping. However, the effect of PTS insertion in pressure transducer kits on the frequency characteristics is minimal. Thus, we found that the use of PTS markedly improved the frequency characteristics as compared to the use of FTP.