Improving water quality in a dialysis unit using root cause analysis.

Background Water quality levels from hemodialysis (HD) and reverse osmosis (RO) machines in dialysis units must meet standards set by the American Association of Medical Instrumentation. Researchers used a root cause analysis (RCA) approach to identify and address factors affecting water quality in...

Descripción completa

Detalles Bibliográficos
Publicado en:American Journal of Infection Control Vol. 45; no. 7; pp. 799 - 805
Autores principales: Yadav, Pallavi, England, Dawn, Vanderkolk, Caprice, Iroh Tam, Pui-Ying
Formato: research Journal Article
Publicado: Elsevier B.V. Jul2017
Acceso en línea:Ver este registro en EBSCOhost
Descripción
Sumario:Background Water quality levels from hemodialysis (HD) and reverse osmosis (RO) machines in dialysis units must meet standards set by the American Association of Medical Instrumentation. Researchers used a root cause analysis (RCA) approach to identify and address factors affecting water quality in the HD and portable RO machines at our institution. Methods A multidisciplinary team reviewed processes, interviewed staff members, and identified opportunities to improve the current sampling and machine disinfection processes. The RCA team identified and implemented 5 interventions, of which 3 were process (changes in water sampling technique, machine disinfection processes, and allocation of machine maintenance duties) and 2 were structural (regular cleaning of water sampling tubes and spigots and addition of new water sampling sites in the system) measures. Results Postimplementation of new protocols, 100% of water cultures of HD and RO machines consistently met the required regulatory standards as recorded over a period of 8 months. Conclusions RCA approach helped improve patient safety, quality of care, streamlined processes, and improved efficiencies of work for staff within the HD program.