Lung deposition and systemic bioavailability of different aerosol devices with and without humidification in mechanically ventilated patients.

Background During mechanical ventilation medical aerosol delivery has been reported to be upto two fold greater with dry inhaled gas than with heated humidity. Urine levels at 0.5 h post dose (URSAL0.5%) has been confirmed as an index of lung deposition and 24 h (URSAL24%) as index of systemic absor...

Descripción completa

Detalles Bibliográficos
Publicado en:Heart & Lung Vol. 46; no. 6; pp. 464 - 468
Autores principales: Moustafa, Islam O.F., Ali, Mohammed R. A.-A., Al Hallag, Moataz, Rabea, Hoda, Fink, James B., Dailey, Patricia, Abdelrahim, Mohamed E.A.
Formato: research Journal Article
Publicado: Elsevier B.V. Nov2017
Acceso en línea:Ver este registro en EBSCOhost
Descripción
Sumario:Background During mechanical ventilation medical aerosol delivery has been reported to be upto two fold greater with dry inhaled gas than with heated humidity. Urine levels at 0.5 h post dose (URSAL0.5%) has been confirmed as an index of lung deposition and 24 h (URSAL24%) as index of systemic absorption. Our aim was to determine the effect of humidification and aerosol device type on drug delivery to ventilated patients using urine levels. Methods In a randomized crossover design, 36 (18female) mechanically ventilated patients were assigned to one of three groups. Groups 1 and 2 received 5000 μg salbutamol using vibrating mesh (VM) and jet nebulizers (JN), respectively, while group 3 received 1600 μg (16 puffs) of salbutamol via metered dose inhaler with AeroChamber Vent (MDI-AV). All devices were placed in the inspiratory limb of ventilator downstream from the humidifier. Each subject received aerosol with and without humidity at >24 h intervals with >12 h washout periods between salbutamol doses. Patients voided urine 15 min before each study dose and urine samples were collected at 0.5 h post dosing and pooled for the next 24 h. Results The MDI-AV and VM resulted in a higher percentage of urinary salbutamol levels compared to the JN (p < 0.05). Urine levels were similar between humidity and dry conditions. Conclusions Our findings suggest that in-vitro reports overestimate the impact of dry vs. heated humidified conditions on the delivery of aerosol during invasive mechanical ventilation.