Avoiding damage to transvenous leads—A comparison of electrocautery techniques and two insulated electrocautery blades.

Background: Electrocautery (cautery) can damage transvenous cardiac device leads. The purpose of this study was to compare lead damage from an insulated cautery blade when used with several different techniques that included coagulation (COAG) versus cutting (CUT) mode, perpendicular active edge (ac...

Full description

Bibliographic Details
Published in:Pacing & Clinical Electrophysiology Vol. 41; no. 12; pp. 1593 - 1600
Main Authors: Wasserlauf, Jeremiah, Esheim, Taiki, Jarett, Natasha M., Chan, Eric K.Y., Schaller, Robert D., Garcia, Fermin C., Knight, Bradley P.
Format: pictorial research tables/charts Journal Article
Published: Wiley-Blackwell Dec2018
Online Access:View this record in EBSCOhost
fields @attributes:
  recordID: 1
pdfLink:
plink: https://search.ebscohost.com/login.aspx?direct=true&db=ccm&AN=133318322&site=ehost-live
header:
  @attributes:
    shortDbName: ccm
    uiTerm: 133318322
    longDbName: CINAHL Complete
    uiTag: AN
  controlInfo:
    bkinfo:
    dissinfo:
    jinfo:
      jid:
        01478389
        4F8
      jtl: Pacing & Clinical Electrophysiology
      issn: 01478389
      maglogo: Y
    pubinfo:
      dt: Dec2018
      vid: 41
      iid: 12
      pid: 480
      pub: Wiley-Blackwell
      place: Malden, Massachusetts
    artinfo:
      ui:
        133318322
        133318322
        133318322
        10.1111/pace.13519
        133318322
      ppf: 1593
      ppct: 7
      formats:
        fmt:
          – @attributes:
              type: T
          – @attributes:
              type: C
          – @attributes:
              type: P
      tig:
        atl: Avoiding damage to transvenous leads—A comparison of electrocautery techniques and two insulated electrocautery blades.
      aug:
        au:
          Wasserlauf, Jeremiah
          Esheim, Taiki
          Jarett, Natasha M.
          Chan, Eric K.Y.
          Schaller, Robert D.
          Garcia, Fermin C.
          Knight, Bradley P.
        affil: Northwestern University, Feinberg School of Medicine, Chicago IL, USA
      sug:
        subj:
          Cardiac Pacing, Artificial
          Electrocoagulation Methods
          Electrocoagulation Equipment and Supplies
          Defibrillators, Implantable
          Prosthesis Failure Prevention and Control
          Human
          Polyurethanes
          Silicones
          Polymers
      ab: Background: Electrocautery (cautery) can damage transvenous cardiac device leads. The purpose of this study was to compare lead damage from an insulated cautery blade when used with several different techniques that included coagulation (COAG) versus cutting (CUT) mode, perpendicular active edge (active) versus parallel flat blade (flat) orientation (phase 1), and using one commercially available blade (PhotonBlade) versus another (PlasmaBlade) (phase 2). Methods: In phase 1, lesions were delivered using combinations of: (1) COAG and CUT; (2) active and flat orientation; and (3) polyurethane, silicone, and copolymer insulation. In phase 2, lesions were delivered using combinations of: (1) PlasmaBlade and PhotonBlade, (2) four power output levels, and (3) eight different lead models. Lead damage was scored on an ordinal scale of 0 to 4. Results: Phase 1: more leads were damaged using COAG than CUT (48% vs 2%, P < 0.0001). When using COAG, 74% of lesions using active orientation had damage versus 22% of lesions using flat orientation (P  =  0.0002). COAG lesions to copolymer (61%) and polyurethane (68%) leads had greater damage than silicone (17%) (P  =  0.006 and P  =  0.003, respectively). Phase 2: 75% of treatments using PlasmaBlade had damage versus 40% of treatments with PhotonBlade (P < 0.0001). Higher power resulted in more damage. At the commonly used setting of CUT 20 W, damage occurred in 39% of treatments using PlasmaBlade versus 13% using PhotonBlade (P  =  0.0006). Conclusions: COAG resulted in more damage than CUT; this effect was greatest with the active edge, and with polyurethane or copolymer insulation. PhotonBlade was associated with less damage to leads than PlasmaBlade.
      pubtype: Academic Journal
      doctype:
        pictorial
        research
        tables/charts
        Journal Article
      ougenre: Article
    language: English
    refInfo:
    holdings:
      @attributes:
        islocal: N