Difference in specimen weights with semi-automatic cutting biopsy needles.

Purpose: To assess specimen weight difference of six types of semi-automatic cutting biopsy needles.Materials and Methods: We compared 18- and 20-gauge needles, one aspiration-type (STARCUT® aspiration-type, TSK Laboratory, Tochigi, Japan) and five non-aspiration-type (MISSION®, BARD, AZ; SuperCore™...

Descripción completa

Detalles Bibliográficos
Publicado en:Japanese Journal of Radiology Vol. 38; no. 6; pp. 579 - 587
Autores principales: Iguchi, Toshihiro, Hiraki, Takao, Matsui, Yusuke, Mitsuhashi, Toshiharu, Komaki, Toshiyuki, Okamoto, Soichiro, Fukuma, Shogo, Tomita, Koji, Uka, Mayu, Kanazawa, Susumu
Formato: Journal Article
Publicado: Springer Nature Jun2020
Acceso en línea:Ver este registro en EBSCOhost
fields @attributes:
  recordID: 1
pdfLink:
plink: https://search.ebscohost.com/login.aspx?direct=true&db=ccm&AN=143396562&site=ehost-live
header:
  @attributes:
    shortDbName: ccm
    uiTerm: 143396562
    longDbName: CINAHL Complete
    uiTag: AN
  controlInfo:
    bkinfo:
    dissinfo:
    jinfo:
      jid:
        18671071
        AUCM
      jtl: Japanese Journal of Radiology
      issn: 18671071
      maglogo: N
    pubinfo:
      dt: Jun2020
      vid: 38
      iid: 6
      pid: 237
      pub: Springer Nature
      place: New York, New York
    artinfo:
      ui:
        143396562
        143396562
        NLM32124166
        10.1007/s11604-020-00935-3
        NLM32124166
        143396562
      ppf: 579
      ppct: 8
      formats:
        fmt:
          – @attributes:
              type: T
          – @attributes:
              type: P
      tig:
        atl: Difference in specimen weights with semi-automatic cutting biopsy needles.
      aug:
        au:
          Iguchi, Toshihiro
          Hiraki, Takao
          Matsui, Yusuke
          Mitsuhashi, Toshiharu
          Komaki, Toshiyuki
          Okamoto, Soichiro
          Fukuma, Shogo
          Tomita, Koji
          Uka, Mayu
          Kanazawa, Susumu
        affil: Department of Radiology, Okayama University Medical School, 2-5-1 Shikata-cho kita-ku, 700-8558, Okayama, Japan
      sug:
        subj:
          Liver Pathology
          Specimen Handling Equipment and Supplies
          Animals
          Female
          Biopsy, Needle Equipment and Supplies
          Equipment Design
          Models, Biological
          Cattle
          Male
          Scales
          Female
          Male
      ab: Purpose: To assess specimen weight difference of six types of semi-automatic cutting biopsy needles.Materials and Methods: We compared 18- and 20-gauge needles, one aspiration-type (STARCUT® aspiration-type, TSK Laboratory, Tochigi, Japan) and five non-aspiration-type (MISSION®, BARD, AZ; SuperCore™, Argon Medical Devices, TX; Temno Evolution®, Care Fusion, IL; FINE CORE®, Toray Medical, Tokyo, Japan; Quick-Core®, Cook, IN) needles. Four biopsies were performed with each needle with the longest throw length on an excised bovine liver. The biopsies were repeated with new needles, four times with four different livers. STARCUT® was used both with and without aspiration.Results: Sixteen specimens were obtained with each needle. In needles of gauges, STARCUT® with aspiration provided the heaviest specimen and significantly heavier specimens were obtained with STARCUT® with aspiration (P < 0.05) than five non-aspiration-type needles. The specimen weight differed significantly (P < 0.001) among all 18- and 20-gauge needles. The specimen weights did not differ significantly between aspiration and non-aspiration biopsies with STARCUT® (6.32 vs. 5.97 mg with 18-gauge needle, P = 0.342; 1.95 vs. 1.92 mg with 20-gauge needle, P = 0.886).Conclusion: Although STARCUT® with aspiration provided the heaviest specimen, specimen weights were not significantly different between aspiration and non-aspiration biopsies. We assessed the specimen weight difference of six types of semi-automatic cutting biopsy needles. Significantly heavier specimens were obtained with STARCUT® with aspiration than the other needles. The specimen weight differed significantly among all 18- and 20-gauge needles but did not differ significantly between aspiration and non-aspiration biopsies with STARCUT®.
      pubtype: Academic Journal
      doctype: Journal Article
      ougenre: Article
    language: English
    refInfo:
    holdings:
      @attributes:
        islocal: N