Adaptive Filtering Improved Apnea Detection Performance Using Tracheal Sounds in Noisy Environment: A Simulation Study.

Objective. Tracheal sounds were used to detect apnea on various occasions. However, ambient noises can contaminate tracheal sounds which result in poor performance of apnea detection. The objective of this paper was to apply the adaptive filtering (AF) algorithm to improve the quality of tracheal so...

Full description

Bibliographic Details
Published in:BioMed Research International pp. 1 - 9
Main Authors: Wu, Yanan, Liu, Jing, He, Baolin, Zhang, Xiaotong, Yu, Lu
Format: equations & formulas pictorial research tables/charts Journal Article
Published: Wiley-Blackwell 5/27/2020
Online Access:View this record in EBSCOhost
fields @attributes:
  recordID: 1
pdfLink:
plink: https://search.ebscohost.com/login.aspx?direct=true&db=ccm&AN=143456063&site=ehost-live
header:
  @attributes:
    shortDbName: ccm
    uiTerm: 143456063
    longDbName: CINAHL Complete
    uiTag: AN
  controlInfo:
    bkinfo:
    dissinfo:
    jinfo:
      jid:
        23146133
        FT2T
      jtl: BioMed Research International
      issn: 23146133
      maglogo: N
    pubinfo:
      dt: 5/27/2020
      pid: 480
      pub: Wiley-Blackwell
      place: Malden, Massachusetts
    artinfo:
      ui:
        143456063
        143456063
        143456063
        10.1155/2020/7429345
        143456063
      ppf: 1
      ppct: 8
      formats:
        fmt:
          – @attributes:
              type: T
          – @attributes:
              type: P
      tig:
        atl: Adaptive Filtering Improved Apnea Detection Performance Using Tracheal Sounds in Noisy Environment: A Simulation Study.
      aug:
        au:
          Wu, Yanan
          Liu, Jing
          He, Baolin
          Zhang, Xiaotong
          Yu, Lu
        affil: Department of Biomedical Engineering, School of Fundamental Sciences, China Medical University, Shenyang, Liaoning, China
      sug:
        subj:
          Algorithms
          Sleep Apnea Syndromes Diagnosis
          Trachea Physiology
          Sound Evaluation
          Noise
          Sensitivity and Specificity
          Human
          Simulations
          Transducers
          Predictive Value of Tests
          Adolescence
          Young Adult
          Adult
          Body Mass Index
          kappa Statistic
          Adolescent: 13-18 years
          Adult: 19-44 years
      ab: Objective. Tracheal sounds were used to detect apnea on various occasions. However, ambient noises can contaminate tracheal sounds which result in poor performance of apnea detection. The objective of this paper was to apply the adaptive filtering (AF) algorithm to improve the quality of tracheal sounds and examine the accuracy of the apnea detection algorithm using tracheal sounds after AF. Method. Tracheal sounds were acquired using a primary microphone encased in a plastic bell, and the ambient noises were collected using a reference microphone resting outside the plastic bell in quiet and noisy environments, respectively. Simultaneously, the flow pressure signals and thoracic and abdominal movement were obtained as the standard signals to determine apnea events. Then, the normalized least mean square (NLMS) AF algorithm was applied to the tracheal sounds mixed with noises. Finally, the algorithm of apnea detection was used to the tracheal sounds with AF and the tracheal sounds without AF. Sensitivity, specificity, positive predictive value (PPV), negative predictive value (NPV), accuracy, and Cohen's kappa coefficient of apnea detection were calculated. Results. Forty-six healthy subjects, aged 18-35 years and with BMI < 21.4 , were included in the study. The apnea detection performance using tracheal sounds was as follows: in the quiet environment, the tracheal sounds without AF detected apnea with 97.2% sensitivity, 99.9% specificity, 99.8% PPV, 99.4% NPV, 99.5% accuracy, and 0.982 kappa coefficient. The tracheal sounds with AF detected apnea with 98.2% sensitivity, 99.9% specificity, 99.4% PPV, 99.6% NPV, 99.6% accuracy, and 0.985 kappa coefficient. While in the noisy environment, the tracheal sounds without AF detected apnea with 81.1% sensitivity, 96.9% specificity, 85.1% PPV, 96% NPV, 94.2% accuracy, and 0.795 kappa coefficient and the tracheal sounds with AF detected apnea with 91.5% sensitivity, 97.4% specificity, 88.4% PPV, 98.2% NPV, 96.4% accuracy, and 0.877 kappa coefficient. Conclusion. The performance of apnea detection using tracheal sounds with the NLMS AF algorithm in the noisy environment proved to be accurate and reliable. The AF technology could be applied to the respiratory monitoring using tracheal sounds.
      pubtype: Academic Journal
      doctype:
        equations & formulas
        pictorial
        research
        tables/charts
        Journal Article
      ougenre: Article