MRI following scoliosis surgery? An analysis of implant heating, displacement, torque, and susceptibility artifacts.

Objectives: The implant constructs used in scoliosis surgery are often long with a high screw density. Therefore, it is generally believed that magnetic resonance imaging (MRI) should not be carried out after scoliosis surgery, with the result that computed tomography is often preferred despite the...

Descripción completa

Detalles Bibliográficos
Publicado en:European Radiology Vol. 31; no. 6; pp. 4298 - 4308
Autores principales: Heinrich, Andreas, Reinhold, Maximilian, Güttler, Felix V., Matziolis, Georg, Teichgräber, Ulf K.-M., Zippelius, Timo, Strube, Patrick
Formato: research tables/charts Journal Article
Publicado: Springer Nature Jun2021
Acceso en línea:Ver este registro en EBSCOhost
fields @attributes:
  recordID: 1
pdfLink:
plink: https://search.ebscohost.com/login.aspx?direct=true&db=ccm&AN=150343656&site=ehost-live
header:
  @attributes:
    shortDbName: ccm
    uiTerm: 150343656
    longDbName: CINAHL Complete
    uiTag: AN
  controlInfo:
    bkinfo:
    dissinfo:
    jinfo:
      jid:
        09387994
        NPH
      jtl: European Radiology
      issn: 09387994
      maglogo: N
    pubinfo:
      dt: Jun2021
      vid: 31
      iid: 6
      pid: 237
      pub: Springer Nature
      place: New York, New York
    artinfo:
      ui:
        150343656
        147391464
        150343656
        NLM33277671
        150343656
        10.1007/s00330-020-07546-6
        NLM33277671
        150343656
      ppf: 4298
      ppct: 10
      formats:
        fmt:
          @attributes:
            type: P
      tig:
        atl: MRI following scoliosis surgery? An analysis of implant heating, displacement, torque, and susceptibility artifacts.
      aug:
        au:
          Heinrich, Andreas
          Reinhold, Maximilian
          Güttler, Felix V.
          Matziolis, Georg
          Teichgräber, Ulf K.-M.
          Zippelius, Timo
          Strube, Patrick
        affil: Department of Radiology, Jena University Hospital – Friedrich Schiller University, Am Klinikum 1, 07747, Jena, Germany
      sug:
        subj:
          Scoliosis
          Scoliosis Surgery
          Artifacts
          Torque
          Magnetic Resonance Imaging
          Heating
          Human
      ab: Objectives: The implant constructs used in scoliosis surgery are often long with a high screw density. Therefore, it is generally believed that magnetic resonance imaging (MRI) should not be carried out after scoliosis surgery, with the result that computed tomography is often preferred despite the ionizing radiation involved. The objective of this study was to evaluate the MRI compatibility of long pedicle-screw-rod constructs at 1.5 T and 3 T using standardized methods of the American Society for Testing and Materials (ASTM).Methods: Constructs between 130 and 430 mm long were systematically examined according to the ASTM standards F2182 (radio frequency-induced heating), F2119 (susceptibility artifacts), F2213 (magnetically induced torque), and F2052 (magnetically induced displacement force).Results: The maximum heating in the magnetic field was 1.3 K. Heating was significantly influenced by magnetic field strength (p < 0.001), implant length (p = 0.048), and presence of cross-links (p = 0.001). The maximum artifact width for different lengths of the anatomically bent titanium rods with CoCr alloy ranged between 14.77 ± 2.93 mm (TSE) and 17.49 ± 1.82 mm (GRE) for 1.5 T and between 23.67 ± 2.39 mm (TSE) and 27.77 ± 2.37 mm (GRE) for 3 T. TiCP and TiAl showed the smallest and CoCr and CoCr Plus the largest artifact widths. The magnetically induced torque and displacement force were negligible.Conclusions: MRI following scoliosis surgery with long implant constructs is safe with the patient in supine position. Although susceptibility artifacts can severely limit the diagnostic value, the examination of other regions is possible.Key Points: • Large spinal implants are not necessarily a contraindication for MRI; MR conditional status can be examined according to the ASTM standards F2182, F2119, F2213, and F2052. • A metallic pedicle-screw-rod system could be reliably and safely examined in all combinations of length (130 to 430 mm), configuration, and material in a B0 at 1.5 T and 3 T. • According to ASTM F2503, the examined pedicle-screw-rod system is MR conditional and especially the young patients can benefit from a non-ionizing radiation MRI examination.
      pubtype: Academic Journal
      doctype:
        research
        tables/charts
        Journal Article
      ougenre: Article
    language: English
    refInfo:
    holdings:
      @attributes:
        islocal: N