Heated Humidifiers for Noninvasive Respiratory Support and the Risk of Burns in Neonates: A Bench Evaluation.

BACKGROUND: User errors in managing heated humidifiers (HHs) have been suggested to be a source of nasal burns in newborns treated with nasal CPAP. This study evaluated the risk of burns by reproducing 3 typical errors concerning the use of HHs. METHODS: Six HHs were tested on a bench in a tradition...

Descripción completa

Detalles Bibliográficos
Publicado en:Respiratory Care Vol. 66; no. 9; pp. 1398 - 1406
Autores principales: Fau, Sebastien, Baud, Olivier, Rimensberger, Peter
Formato: CEU pictorial research tables/charts Journal Article
Publicado: Mary Ann Liebert, Inc. Sep2021
Acceso en línea:Ver este registro en EBSCOhost
fields @attributes:
  recordID: 1
pdfLink:
plink: https://search.ebscohost.com/login.aspx?direct=true&db=ccm&AN=152131913&site=ehost-live
header:
  @attributes:
    shortDbName: ccm
    uiTerm: 152131913
    longDbName: CINAHL Complete
    uiTag: AN
  controlInfo:
    bkinfo:
    dissinfo:
    jinfo:
      jid:
        00201324
        4GG
      jtl: Respiratory Care
      issn: 00201324
      maglogo: N
    pubinfo:
      dt: Sep2021
      vid: 66
      iid: 9
      pid: 1365
      pub: Mary Ann Liebert, Inc.
      place: New Rochelle, New York
    artinfo:
      ui:
        152131913
        152131913
        152131913
        10.4187/respcare.08768
        152131913
      ppf: 1398
      ppct: 8
      formats:
        fmt:
          @attributes:
            type: P
      tig:
        atl: Heated Humidifiers for Noninvasive Respiratory Support and the Risk of Burns in Neonates: A Bench Evaluation.
      aug:
        au:
          Fau, Sebastien
          Baud, Olivier
          Rimensberger, Peter
        affil: Service of Neonatology and Pediatric Intensive Care, Department of Pediatrics, University Hospital of Geneva, Geneva, Switzerland
      sug:
        subj:
          Nebulizers and Vaporizers Adverse Effects
          Heat Adverse Effects
          Continuous Positive Airway Pressure
          Burns Risk Factors
          Risk Assessment
          Human Error Evaluation
          Education, Continuing (Credit)
          Human
          Infant, Newborn
          Temperature Evaluation
          Scales
          Descriptive Statistics
          Academic Medical Centers
          Switzerland
          Data Analysis Software
          Patient Safety
          Algorithms
          Iatrogenic Disease
          Infant, Newborn: birth-1 month
      ab: BACKGROUND: User errors in managing heated humidifiers (HHs) have been suggested to be a source of nasal burns in newborns treated with nasal CPAP. This study evaluated the risk of burns by reproducing 3 typical errors concerning the use of HHs. METHODS: Six HHs were tested on a bench in a traditional nasal CPAP setup: PMH5000, Aircon (Wilamed); MR730, MR850, MR950 (Fisher & Paykel); and H900 (Hamilton). Temperature was measured at the end of the inspiratory tubing limb. Errors tested were (1) misconnection of the HH thermal probes (NoProbe), (2) absence of gas flow while the HH is on (NoFlow), and (3) unsuitable repeated acknowledgment of the HH alarm (NoAlarm). These errors were combined in 3 standardized scenarios: (1) NoProbe + NoFlow + NoAlarm; (2) NoProbe + NoAlarm, and (3) NoFlow + NoAlarm. The NoProbe + NoFlow + NoAlarm and NoProbe + NoAlarm scenarios were not tested in the H900 and MR950 because the proprietary circuits of these HHs are equipped with embedded probes. RESULTS: For each HH, the highest inspiratory gas temperature (HIGT) and the rating on a self-designed risk-of-burn scale (ie, no risk, moderate risk, or severe risk) were reported. In the NoProbe + NoFlow + NoAlarm scenario, the risk was severe for the MR730, PMH5000, MR850, and Aircon, with HIGTs of > 65°C, 58°C, 56°C, and > 65°C, respectively. In the NoProbe + NoAlarm scenario, the risk was also severe for the same 4 HHs, with HIGTs of 56°C, 47°C, 56°C, and 48°C, respectively. In the NoFlow + NoAlarm scenario, the risk was severe for the PMH5000, Aircon, and H900, with HIGTs of 52°C, > 65°C, and 49°C, respectively, and moderate for the MR730, MR850, and MR950, with HIGTs of 45°C, 47°C, and 44°C, respectively). CONCLUSIONS: In case of misuse, 5 of the 6 tested devices presented a severe risk of inducing skin burns, whereas the MR950 presented a moderate risk.
      pubtype: Academic Journal
      doctype:
        CEU
        pictorial
        research
        tables/charts
        Journal Article
      ougenre: Article
    language: English
    refInfo:
    holdings:
      @attributes:
        islocal: N