Inappropriate rate response in a leadless pacemaker due to automatic rate profile optimization.

A three‐axis accelerometer sensor is incorporated in a leadless pacemaker (MicraTM, MedtronicInc) for rate adaptation. Three sensor setpoints at lower rate (LR), activity of daily living rate (ADLR), and upper rate (UR) are used to convert detected acceleration to a desired rate response. We describ...

Descripción completa

Detalles Bibliográficos
Publicado en:Pacing & Clinical Electrophysiology Vol. 45; no. 1; pp. 141 - 145
Autores principales: Lau, Chu‐Pak, Lee, Kathy Lai‐Fun, Tse, Hung‐Fat
Formato: case study pictorial tables/charts Journal Article
Publicado: Wiley-Blackwell Jan2022
Acceso en línea:Ver este registro en EBSCOhost
fields @attributes:
  recordID: 1
pdfLink:
plink: https://search.ebscohost.com/login.aspx?direct=true&db=ccm&AN=154741105&site=ehost-live
header:
  @attributes:
    shortDbName: ccm
    uiTerm: 154741105
    longDbName: CINAHL Complete
    uiTag: AN
  controlInfo:
    bkinfo:
    dissinfo:
    jinfo:
      jid:
        01478389
        4F8
      jtl: Pacing & Clinical Electrophysiology
      issn: 01478389
      maglogo: Y
    pubinfo:
      dt: Jan2022
      vid: 45
      iid: 1
      pid: 480
      pub: Wiley-Blackwell
      place: Malden, Massachusetts
    artinfo:
      ui:
        154741105
        153615002
        154741105
        154741105
        10.1111/pace.14393
        154741105
      ppf: 141
      ppct: 4
      formats:
        fmt:
          – @attributes:
              type: T
          – @attributes:
              type: C
          – @attributes:
              type: P
      tig:
        atl: Inappropriate rate response in a leadless pacemaker due to automatic rate profile optimization.
      aug:
        au:
          Lau, Chu‐Pak
          Lee, Kathy Lai‐Fun
          Tse, Hung‐Fat
        affil: Cardiology Division, Department of Medicine, Queen Mary Hospital, The University of Hong Kong, Hong Kong SAR, China
      sug:
        subj:
          Accelerometers
          Pacemaker, Artificial
          Activities of Daily Living
          Chest Physical Therapy
          Sedentary Behavior In Old Age
          Atrial Fibrillation
          Heart Rate
          Tachycardia
          Aged
          Aged: 65+ years
      ab: A three‐axis accelerometer sensor is incorporated in a leadless pacemaker (MicraTM, MedtronicInc) for rate adaptation. Three sensor setpoints at lower rate (LR), activity of daily living rate (ADLR), and upper rate (UR) are used to convert detected acceleration to a desired rate response. We describe inappropriate rate acceleration at rest and during chest physiotherapy in a sedentary elderly patient with leadless pacemaker. The underlying mechanism and various programming options are discussed.
      pubtype: Academic Journal
      doctype:
        case study
        pictorial
        tables/charts
        Journal Article
      ougenre: Article
    language: English
    refInfo:
    holdings:
      @attributes:
        islocal: N