Hybrid total-body pet scanners—current status and future perspectives.

Purpose Since the 1990s, PET has been successfully combined with MR or CT systems. In the past years, especially PET systems have seen a trend towards an enlarged axial field of view (FOV), up to a factor of ten. Methods Conducting a thorough literature research, we summarize the status quo of conte...

Full description

Bibliographic Details
Published in:European Journal of Nuclear Medicine & Molecular Imaging Vol. 49; no. 2; pp. 445 - 460
Main Authors: Nadig, Vanessa, Herrmann, Ken, Mottaghy, Felix M., Schulz, Volkmar
Format: pictorial review tables/charts Journal Article
Published: Springer Nature Jan2022
Online Access:View this record in EBSCOhost
Description
Summary:Purpose Since the 1990s, PET has been successfully combined with MR or CT systems. In the past years, especially PET systems have seen a trend towards an enlarged axial field of view (FOV), up to a factor of ten. Methods Conducting a thorough literature research, we summarize the status quo of contemporary total-body (TB) PET/CT scanners and give an outlook on possible future developments. Results Currently, three human TB PET/CT systems have been developed: The PennPET Explorer, the uExplorer, and the Biograph Vision Quadra realize aFOVs between 1 and 2 m and show a tremendous increase in system sensitivity related to their longer gantries. Conclusion The increased system sensitivity paves the way for short-term, low-dose, and dynamic TB imaging as well as new examination methods in almost all areas of imaging.