New Software for DQE Calculation in Digital Mammography Compliant with IEC 62220–1-2.

Significant improvements in mammography systems have been achieved with the introduction of active matrix flat-panel digital detectors. The advent of this technology also makes it possible to implement computational methods for quantitative image analysis. This study describes new software created t...

Descripción completa

Detalles Bibliográficos
Publicado en:Journal of Digital Imaging Vol. 35; no. 5; pp. 1069 - 1079
Autores principales: Dobrzyńska, Magdalena, Wysocka-Rabin, Anna, Fabiszewska, Ewa, Pasicz, Katarzyna, Skrzyński, Witold
Formato: equations & formulas pictorial research tables/charts Journal Article
Publicado: Springer Nature Oct2022
Acceso en línea:Ver este registro en EBSCOhost
fields @attributes:
  recordID: 1
pdfLink:
plink: https://search.ebscohost.com/login.aspx?direct=true&db=ccm&AN=159758917&site=ehost-live
header:
  @attributes:
    shortDbName: ccm
    uiTerm: 159758917
    longDbName: CINAHL Complete
    uiTag: AN
  controlInfo:
    bkinfo:
    dissinfo:
    jinfo:
      jid:
        08971889
        DOQ
      jtl: Journal of Digital Imaging
      issn: 08971889
      maglogo: N
    pubinfo:
      dt: Oct2022
      vid: 35
      iid: 5
      pid: 237
      pub: Springer Nature
      place: New York, New York
    artinfo:
      ui:
        159758917
        159099403
        159758917
        159758917
        10.1007/s10278-021-00546-y
        159758917
      ppf: 1069
      ppct: 10
      formats:
        fmt:
          – @attributes:
              type: T
          – @attributes:
              type: P
      tig:
        atl: New Software for DQE Calculation in Digital Mammography Compliant with IEC 62220–1-2.
      aug:
        au:
          Dobrzyńska, Magdalena
          Wysocka-Rabin, Anna
          Fabiszewska, Ewa
          Pasicz, Katarzyna
          Skrzyński, Witold
        affil: Particle Acceleration Physics and Technology Division, National Centre for Nuclear Research (NCNR), Otwock Swierk, Poland
      sug:
        subj:
          Mammography Methods
          Software Design Evaluation
          Noise
          User-Computer Interface
          Human
          Algorithms
          Uncertainty
          Image Processing, Computer Assisted Methods
          Practice Guidelines
      ab: Significant improvements in mammography systems have been achieved with the introduction of active matrix flat-panel digital detectors. The advent of this technology also makes it possible to implement computational methods for quantitative image analysis. This study describes new software created to perform detective quantum efficiency (DQE) calculations fully compliant with the IEC 62220–1-2 standard. Python-based software was developed that contains modules to calculate inverse conversion function, modulation transfer function (MTF), noise power spectrum (NPS), and DQE itself. A graphical user interface (GUI) and further add-ons make this software more user-friendly. Results are immediately displayed diagrammatically, and complete output data are exported to a.csv file. The code is available freely, as a compiled, executable file (.exe). The program was successfully tested using DICOM images obtained from mammography units from different manufacturers. This study also includes validation of the new software, based on comparisons of results obtained for the same set of data with two other, freely available programs.