Comparison of liver iron concentration calculated from R2* at 1.5 T and 3 T.

Purpose: R2*, a measurement obtained using magnetic resonance imaging (MRI) can be used to estimate liver iron concentration (LIC). 3 T and 1.5 T scanners can be used but conversion of 3 T R2* to LIC is less well validated. In this study the aim was to compare 3 T-R2* LIC and 1.5 T-R2* LIC estimatio...

Descripción completa

Detalles Bibliográficos
Publicado en:Abdominal Radiology Vol. 48; no. 3; pp. 865 - 874
Autores principales: Pickles, Elisabeth, Kumar, Shravan, Brady, Michael, Telford, Alison, Pavlides, Michael, Bulte, Daniel
Formato: Journal Article
Publicado: Springer Nature Mar2023
Acceso en línea:Ver este registro en EBSCOhost
fields @attributes:
  recordID: 1
pdfLink:
plink: https://search.ebscohost.com/login.aspx?direct=true&db=ccm&AN=161991646&site=ehost-live
header:
  @attributes:
    shortDbName: ccm
    uiTerm: 161991646
    longDbName: CINAHL Complete
    uiTag: AN
  controlInfo:
    bkinfo:
    dissinfo:
    jinfo:
      jid:
        2366004X
        JT14
      jtl: Abdominal Radiology
      issn: 2366004X
      maglogo: N
    pubinfo:
      dt: Mar2023
      vid: 48
      iid: 3
      pid: 237
      pub: Springer Nature
      place: New York, New York
    artinfo:
      ui:
        161991646
        160800070
        10.1007/s00261-022-03762-4
        161991646
      ppf: 865
      ppct: 9
      formats:
        fmt:
          – @attributes:
              type: T
          – @attributes:
              type: P
      tig:
        atl: Comparison of liver iron concentration calculated from R2* at 1.5 T and 3 T.
      aug:
        au:
          Pickles, Elisabeth
          Kumar, Shravan
          Brady, Michael
          Telford, Alison
          Pavlides, Michael
          Bulte, Daniel
        affil: Institute of Biomedical Engineering, Department of Engineering Science, University of Oxford, Old Road Campus Research Building, OX3 7DQ, Headington, Oxford, UK
      sug:
      ab: Purpose: R2*, a measurement obtained using magnetic resonance imaging (MRI) can be used to estimate liver iron concentration (LIC). 3 T and 1.5 T scanners can be used but conversion of 3 T R2* to LIC is less well validated. In this study the aim was to compare 3 T-R2* LIC and 1.5 T-R2* LIC estimations to assess if they can be used interchangeably. Methods: Thirty participants were scanned at both 1.5 T and 3 T. R2* was measured at both field strengths. 3 T R2* and 1.5 R2* were compared using linear regression and were converted to LIC using different calibration curves. Pearson's rho and Intraclass Correlation Coefficients (ICCs) were used to assess correlation and agreement between 1.5 and 3 T LIC. Bland Altman plots were used to assess bias and limits of agreement. Results: All 1.5 T and 3 T LIC comparisons gave Pearson's rho of 0.99 (p < 0.001). ICC ranged from 0.83 (p = 0.005) to 0.96 (p < 0.001). Biases had magnitude of less than 0.2 mg/g dry weight. Conclusion: Agreement and bias between 3 and 1.5 T-R2* LIC depended on the method used for conversion. There were instances when the agreement was excellent and bias was small, indicating that potentially 3 T-R2* LIC can be used alongside or instead of 1.5 T-R2* LIC but care needs to be taken over the conversion methods selected. Trial registration number: Clinicaltrials.gov NCT03743272, 16 November 2018.
      pubtype: Academic Journal
      doctype: Journal Article
      ougenre: Article
    language: English
    refInfo:
    holdings:
      @attributes:
        islocal: N