Development and performance of SIAT bPET: a high-resolution and high-sensitivity MR-compatible brain PET scanner using dual-ended readout detectors.

Purpose: Positron emission tomography/magnetic resonance imaging (PET/MRI) is a powerful tool for brain imaging, but the spatial resolution of the PET scanners currently used for brain imaging can be further improved to enhance the quantitative accuracy of brain PET imaging. The purpose of this stud...

Descripción completa

Detalles Bibliográficos
Publicado en:European Journal of Nuclear Medicine & Molecular Imaging Vol. 51; no. 2; pp. 346 - 358
Autores principales: Kuang, Zhonghua, Sang, Ziru, Ren, Ning, Wang, Xiaohui, Zeng, Tianyi, Wu, San, Niu, Ming, Cong, Longhan, Kinyanjui, Samuel M., Chen, Qiaoyan, Tie, Changjun, Liu, Zheng, Sun, Tao, Hu, Zhanli, Du, Junwei, Li, Ye, Liang, Dong, Liu, Xin, Zheng, Hairong, Yang, Yongfeng
Formato: Journal Article
Publicado: Springer Nature Jan2024
Acceso en línea:Ver este registro en EBSCOhost
fields @attributes:
  recordID: 1
pdfLink:
plink: https://search.ebscohost.com/login.aspx?direct=true&db=ccm&AN=174658112&site=ehost-live
header:
  @attributes:
    shortDbName: ccm
    uiTerm: 174658112
    longDbName: CINAHL Complete
    uiTag: AN
  controlInfo:
    bkinfo:
    dissinfo:
    jinfo:
      jid:
        16197070
        NPC
      jtl: European Journal of Nuclear Medicine & Molecular Imaging
      issn: 16197070
      maglogo: N
    pubinfo:
      dt: Jan2024
      vid: 51
      iid: 2
      pid: 237
      pub: Springer Nature
      place: New York, New York
    artinfo:
      ui:
        174658112
        173358884
        10.1007/s00259-023-06458-z
        174658112
      ppf: 346
      ppct: 12
      formats:
        fmt:
          – @attributes:
              type: T
          – @attributes:
              type: P
      tig:
        atl: Development and performance of SIAT bPET: a high-resolution and high-sensitivity MR-compatible brain PET scanner using dual-ended readout detectors.
      aug:
        au:
          Kuang, Zhonghua
          Sang, Ziru
          Ren, Ning
          Wang, Xiaohui
          Zeng, Tianyi
          Wu, San
          Niu, Ming
          Cong, Longhan
          Kinyanjui, Samuel M.
          Chen, Qiaoyan
          Tie, Changjun
          Liu, Zheng
          Sun, Tao
          Hu, Zhanli
          Du, Junwei
          Li, Ye
          Liang, Dong
          Liu, Xin
          Zheng, Hairong
          Yang, Yongfeng
        affil: Shenzhen Institute of Advanced Technology, Chinese Academy of Sciences, 518055, Shenzhen, China
      sug:
      ab: Purpose: Positron emission tomography/magnetic resonance imaging (PET/MRI) is a powerful tool for brain imaging, but the spatial resolution of the PET scanners currently used for brain imaging can be further improved to enhance the quantitative accuracy of brain PET imaging. The purpose of this study is to develop an MR-compatible brain PET scanner that can simultaneously achieve a uniform high spatial resolution and high sensitivity by using dual-ended readout depth encoding detectors. Methods: The MR-compatible brain PET scanner, named SIAT bPET, consists of 224 dual-ended readout detectors. Each detector contains a 26 × 26 lutetium yttrium oxyorthosilicate (LYSO) crystal array of 1.4 × 1.4 × 20 mm3 crystal size read out by two 10 × 10 silicon photomultiplier (SiPM) arrays from both ends. The scanner has a detector ring diameter of 376.8 mm and an axial field of view (FOV) of 329 mm. The performance of the scanner including spatial resolution, sensitivity, count rate, scatter fraction, and image quality was measured. Imaging studies of phantoms and the brain of a volunteer were performed. The mutual interferences of the PET insert and the uMR790 3 T MRI scanner were measured, and simultaneous PET/MRI imaging of the brain of a volunteer was performed. Results: A spatial resolution of better than 1.5 mm with an average of 1.2 mm within the whole FOV was obtained. A sensitivity of 11.0% was achieved at the center FOV for an energy window of 350–750 keV. Except for the dedicated RF coil, which caused a ~ 30% reduction of the sensitivity of the PET scanner, the MRI sequences running had a negligible effect on the performance of the PET scanner. The reduction of the SNR and homogeneity of the MRI images was less than 2% as the PET scanner was inserted to the MRI scanner and powered-on. High quality PET and MRI images of a human brain were obtained from simultaneous PET/MRI scans. Conclusion: The SIAT bPET scanner achieved a spatial resolution and sensitivity better than all MR-compatible brain PET scanners developed up to date. It can be used either as a standalone brain PET scanner or a PET insert placed inside a commercial whole-body MRI scanner to perform simultaneous PET/MRI imaging.
      pubtype: Academic Journal
      doctype: Journal Article
      ougenre: Article
    language: English
    refInfo:
    holdings:
      @attributes:
        islocal: N