Effect of a Pedicle Screw Fixation System on Lumbar Spinal Segments: A Finite Element Study.

Objective: Spinal implants have been used to stimulate fusion by surgical adjustment and correct abnormal alignment of the vertebral column. Spinal fusion can cause some spinal disorders and hence describing the changes in biomechanical forces would help to understand these complications. In this st...

Descripción completa

Detalles Bibliográficos
Publicado en:Turkiye Klinikleri Journal of Medical Sciences Vol. 43; no. 4; pp. 372 - 382
Autores principales: AKINCI, Saliha Zeyneb, KARABULUT, Derya, DOĞRU, Suzan Cansel, SÜRMEN, Hasan Kemal, YAMAN, Onur, ARSLAN, Yunus Ziya
Formato: diagnostic images pictorial research tables/charts Journal Article
Publicado: Turkiye Klinikleri ara2023
Acceso en línea:Ver este registro en EBSCOhost
fields @attributes:
  recordID: 1
pdfLink:
plink: https://search.ebscohost.com/login.aspx?direct=true&db=ccm&AN=175291159&site=ehost-live
header:
  @attributes:
    shortDbName: ccm
    uiTerm: 175291159
    longDbName: CINAHL Complete
    uiTag: AN
  controlInfo:
    bkinfo:
    dissinfo:
    jinfo:
      jid:
        13000292
        1VOP
      jtl: Turkiye Klinikleri Journal of Medical Sciences
      issn: 13000292
      maglogo: N
    pubinfo:
      dt: ara2023
      vid: 43
      iid: 4
      pid: 67298
      pub: Turkiye Klinikleri
    artinfo:
      ui:
        175291159
        175291159
        175291159
        10.5336/medsci.2023-99005
        175291159
      ppf: 372
      ppct: 10
      formats:
        fmt:
          @attributes:
            type: P
      tig:
        atl: Effect of a Pedicle Screw Fixation System on Lumbar Spinal Segments: A Finite Element Study.
      aug:
        au:
          AKINCI, Saliha Zeyneb
          KARABULUT, Derya
          DOĞRU, Suzan Cansel
          SÜRMEN, Hasan Kemal
          YAMAN, Onur
          ARSLAN, Yunus Ziya
        affil: Department of Biomedical Engineering and Bioinformatics, İstanbul Medipol University Faculty of Engineering and Natural Sciences, İstanbul, Türkiye
      sug:
        subj:
          Lumbar Vertebrae Surgery
          Pedicle Screws
          Spinal Fusion Adverse Effects
          Postoperative Complications Prevention and Control
          Finite Element Analysis
          Lumbar Vertebrae Physiology
          Biomechanics
          Human
          Treatment Outcomes
          Postoperative Complications Risk Factors
          Models, Anatomic
          Scoliosis Radiography
          Tomography, X-Ray Computed
          Zygapophyseal Joint
          Intervertebral Disk
          Ligaments
          Stress, Mechanical Evaluation
          Bending
          Lumbar Vertebrae Abnormalities
          Bone Malalignment Surgery
          Surgical Patients
          Functional Status
      ab:
        Objective: Spinal implants have been used to stimulate fusion by surgical adjustment and correct abnormal alignment of the vertebral column. Spinal fusion can cause some spinal disorders and hence describing the changes in biomechanical forces would help to understand these complications. In this study, we used two lumbar models. One of them is used without the fixed pedicle screw system, and the other one was used with that system. Therefore, we aimed to investigate the biomechanical effect of a pedicle screw fixation system on the lumbar functional spinal unit under applied forces. Material and Methods: Computed tomography data of a scoliotic patient was used for the construction of the lumbar models. The second and third vertebrae (L2- L3) of the lumbar spine, two facet joints, an intervertebral disc, and ligaments were constructed. A screw fixation system was employed and Von-Mises stress analysis was carried out for both models. Results: The von Mises stress distribution results showed that the presence of fixed implantation transmitted the compressive forces to the screws and rods in all directions and decreased the stress levels considerably by allowing to stabilize the model. The upper side of the L2 vertebra was the most affected region in flexion and lateral bending. However, the pedicle region had the maximum affected area under applied loads in extension and axial rotation. Conclusion: It was concluded that a fixed implant system preserves the maintenance of the vertebral column and decreases the stress on the adjacent spinal segments, especially for the intervertebral discs.
        Amaç: Omurga implantları, cerrahi müdahale ile füzyonu hızlandırmaları ve omurganın anormal eğriliğini düzeltmek için kullanılmaktadır. Spinal füzyon bazı spinal bozukluklara neden olabilir ve bu nedenle füzyon ile lomber omurların komşu bölümlerine uygulanan biyomekanik kuvvetlerdeki değişiklikleri açıklamak, bu komplikasyonları anlamaya yardımcı olacaktır. Bu çalışmada iki tane lomber model kullanıldı. Bir tanesi sabit pedikül vida sistemi olmadan kullanıldı, diğeri ise bu sistem ile birlikte kullanıldı. Bu sebeple, bu çalışmada lomber fonksiyonel omurga biriminde pedikül vida sisteminin uygulanan kuvvetler altındaki biyomekanik etkisini araştırmayı amaçladık. Gereç ve Yöntemler: Skolyoz bir hastanın bilgisayarlı tomografi verileri lomber omurganın temel birimini oluşturmak için kullanılmıştır. Lomber omurganın ikinci ve üçüncü omurları (L2-L3), iki tane faset eklem, bir tane omurlar arası disk ve bağ dokuları oluşturulmuştur. Bir titanyum vida sabitleme sistemi tasarlanmış ve Von Mises gerilme analizleri intakt (implant olmayan sistemler) ve implant olan sistemler için uygulanmıştır. Bulgular: Von Mises gerilme dağılımı sonuçları, sabit implantasyonun olması durumunda baskı yapan kuvvetlerin vidalara ve çubuklara tüm yönlerde iletildiğini göstermiştir ve gerilme seviyelerini önemli ölçüde azaltarak modelin dengede durmasına olanak sağlamıştır. L2 omurunun en üst bölümünün fleksiyon ve yana eğilmede en fazla etkilenen bölge olmuştur. Ancak uygulanan yükler altındaki ekstansiyon ve aksiyal rotasyonda pedikül bölümü en çok etkilenen alan olmuştur. Sonuç: Sabit bir implant sisteminin omurganın bütünlüğünü koruduğu ve özellikle omurlar arası diskler için komşu bölümler üzerindeki gerilmeyi azalttığı sonucuna varılmıştır.
      pubtype: Academic Journal
      doctype:
        diagnostic images
        pictorial
        research
        tables/charts
        Journal Article
      ougenre: Article
    language: English
    refInfo:
    holdings:
      @attributes:
        islocal: N