Importance of dosing: Analysis of touch-free hand hygiene dispensers for consistency.

We investigated 5 touch-free automatic foam alcohol-based handrub (ABHR) hand hygiene dispensers common in US hospitals. Output dosing was evaluated for consistency to comply with the Leapfrog 2022 guidance requirement of ≥ 1.0 mL per ABHR dispense. Fifteen dispensers and 32 distinct refills were te...

Descripción completa

Detalles Bibliográficos
Publicado en:American Journal of Infection Control Vol. 53; no. 6; pp. 696 - 701
Autores principales: Smith, Nicole M., Bánsághi, Száva, Chen, Nanshan, Neal, Travis B., McNulty, John J., Haidegger, Tamás P., Arbogast, James W.
Formato: research Journal Article
Publicado: Elsevier B.V. Jun2025
Acceso en línea:Ver este registro en EBSCOhost