A pre-clinical application study of the SATA-LRS laparoscopic instrument in a human cadaver model.

Background: To validate whether the SATA-LRS, a novel reusable articulating laparoscopic instrument, fits surgical practice, a pre-clinical study was performed. Methods: Thirteen medical doctors used the instrument in a laparoscopic endoscopic inguinal hernia repair (TEP)-like task inside a cadaver....

Descripción completa

Detalles Bibliográficos
Publicado en:Minimally Invasive Therapy & Allied Technologies Vol. 34; no. 6; pp. 447 - 456
Autores principales: Lenssen, Tomas, Postema, Roelf R., Camenzuli, Christian, Calleja-Agius, Jean, Dankelman, Jenny, Horeman-Franse, Tim
Formato: pictorial research tables/charts Journal Article
Publicado: Taylor & Francis Ltd Dec2025
Acceso en línea:Ver este registro en EBSCOhost
fields @attributes:
  recordID: 1
pdfLink:
plink: https://search.ebscohost.com/login.aspx?direct=true&db=ccm&AN=189706709&site=ehost-live
header:
  @attributes:
    shortDbName: ccm
    uiTerm: 189706709
    longDbName: CINAHL Complete
    uiTag: AN
  controlInfo:
    bkinfo:
    dissinfo:
    jinfo:
      jid:
        13645706
        J4S
      jtl: Minimally Invasive Therapy & Allied Technologies
      issn: 13645706
      maglogo: Y
    pubinfo:
      dt: Dec2025
      vid: 34
      iid: 6
      pid: 377
      pub: Taylor & Francis Ltd
      place: Philadelphia, Pennsylvania
    artinfo:
      ui:
        189706709
        188039468
        189706709
        189706709
        10.1080/13645706.2025.2554077
        189706709
      ppf: 447
      ppct: 9
      formats:
      tig:
        atl: A pre-clinical application study of the SATA-LRS laparoscopic instrument in a human cadaver model.
      aug:
        au:
          Lenssen, Tomas
          Postema, Roelf R.
          Camenzuli, Christian
          Calleja-Agius, Jean
          Dankelman, Jenny
          Horeman-Franse, Tim
        affil: Department of Biomechanical Engineering, Delft University of Technology, Delft, The Netherlands
      sug:
        subj:
          Hernia, Inguinal Surgery
          Herniorrhaphy Equipment and Supplies
          Laparoscopy Equipment and Supplies
          Surgical Instruments Evaluation
          Cadaver
          Models, Anatomic
          Robotic Surgical Procedures
          Equipment Design
          Task Performance and Analysis
          Product Evaluation
          Human
          Descriptive Statistics
          T-Tests
          Paired T-Tests
          Time
          Biomechanics
          Funding Source
      ab: Background: To validate whether the SATA-LRS, a novel reusable articulating laparoscopic instrument, fits surgical practice, a pre-clinical study was performed. Methods: Thirteen medical doctors used the instrument in a laparoscopic endoscopic inguinal hernia repair (TEP)-like task inside a cadaver. A set of sensors on the instrument handle detected motion and articulation of the instrument tip. Data from the sensors and video recordings were used to assess the amount and type of movement of the instrument and the time spent on tasks. questionnaire was used to gain insight into the participants' perception of the contextual factors. Results: There was no difference between task time and instrument tip velocities when using articulation (or not) and all participants used articulation at least half of the task time. Instrument-handle movement, indicating the user's hand and arm movement, was significantly reduced when using articulation. The questionnaire indicated strong acceptance of the instrument and the experimental setup, and a desire to use the instrument in surgery by most participants. Conclusions: The added articulation feature of the SATA-LRS instrument was deemed beneficial by the participants, showed no increased handling complexity or time spent on the task and was used frequently when enabled, indicating intuitiveness.
      pubtype: Academic Journal
      doctype:
        pictorial
        research
        tables/charts
        Journal Article
      ougenre: Article
    language: English
    refInfo:
    holdings:
      @attributes:
        islocal: N