The Effect of Independent Humidity Control on Aerosol Drug Delivery During Adult Invasive Mechanical Ventilation.

Background: Aerosol therapy is commonly performed during invasive mechanical ventilation. Although heated humidification is standard practice, studies suggest that humidity can reduce the delivered dose. In this study, we aimed to investigate whether the delivered dose can be increased by decoupling...

Descripción completa

Detalles Bibliográficos
Publicado en:Respiratory Care Vol. 71; no. 8; pp. 827 - 837
Autores principales: Zhang, Madeline X., Huijgen, Tom, Öhrneman, Hugo, Bonn, Daniel, van Rijn, Cees J.M., Löndahl, Jakob
Formato: CEU pictorial research tables/charts Journal Article
Publicado: Mary Ann Liebert, Inc. Aug2026
Acceso en línea:Ver este registro en EBSCOhost
fields @attributes:
  recordID: 1
pdfLink:
plink: https://search.ebscohost.com/login.aspx?direct=true&db=ccm&AN=195834030&site=ehost-live
header:
  @attributes:
    shortDbName: ccm
    uiTerm: 195834030
    longDbName: CINAHL Complete
    uiTag: AN
  controlInfo:
    bkinfo:
    dissinfo:
    jinfo:
      jid:
        00201324
        4GG
      jtl: Respiratory Care
      issn: 00201324
      maglogo: N
    pubinfo:
      dt: Aug2026
      vid: 71
      iid: 8
      pid: 1365
      pub: Mary Ann Liebert, Inc.
      place: New Rochelle, New York
    artinfo:
      ui:
        195834030
        195834030
        195834030
        10.1177/19433654261428090
        195834030
      ppf: 827
      ppct: 10
      formats:
      tig:
        atl: The Effect of Independent Humidity Control on Aerosol Drug Delivery During Adult Invasive Mechanical Ventilation.
      aug:
        au:
          Zhang, Madeline X.
          Huijgen, Tom
          Öhrneman, Hugo
          Bonn, Daniel
          van Rijn, Cees J.M.
          Löndahl, Jakob
        affil: Ms. Zhang, Dr. Bonn, and Dr. van Rijn are affiliated with the Institute of Physics, Van der Waals-Zeeman Institute, University of Amsterdam, Amsterdam, The Netherlands.
      sug:
        subj:
          Aerosols
          Nebulizers and Vaporizers In Adulthood
          Drug Delivery Systems
          Invasive Procedures In Adulthood
          Respiration, Artificial
          Administration, Inhalation
          Education, Continuing (Credit)
          In Vitro Studies
          Male
          Female
          Adult
          Comparative Studies
          Kruskal-Wallis Test
          Mann-Whitney U Test
          Descriptive Statistics
          Confidence Intervals
          Respiratory Therapy Equipment and Supplies
          Models, Biological
          Albuterol Administration and Dosage
          Spectrum Analysis
          Ventilator Patients
          Equipment Design
          Funding Source
          Adult: 19-44 years
          Male
          Female
      ab: Background: Aerosol therapy is commonly performed during invasive mechanical ventilation. Although heated humidification is standard practice, studies suggest that humidity can reduce the delivered dose. In this study, we aimed to investigate whether the delivered dose can be increased by decoupling humidity control from heating during aerosol drug delivery. Methods: In a bench study setup, albuterol sulfate solution was nebulized into an adult invasive mechanical ventilation circuit with a vibrating-mesh nebulizer. An absolute humidifier with decoupled heating and humidification was tested at 0 and 25 mg H2O/L of added humidity during nebulization and compared with 2 conventional pass-over humidifiers with no independent humidity control. The delivered dose, experimentally determined by the drug deposition on a filter between the endotracheal tube and the test lung, was quantified with a UV–Vis spectrometer. The particle size distribution of the aerosol entering the test lung was measured and used to model the lung deposited dose for healthy and diseased lungs. Results: The absolute humidifier at 0 mg/L added humidity led to a significantly higher delivered dose (37.2%) than the pass-over humidifiers (21.2% and 24.1%, P =.01 for both), whereas the absolute humidifier at 25 mg H2O/L added humidity led to a comparable dose (24.9%) as one of the pass-over humidifiers. All these test scenarios achieved sufficient humidity (>33 mg H2O/L) in the gas flow during nebulization. Regional lung deposition simulations suggest that <10% of the nebulized dose deposits in healthy lungs for the pass-over humidifiers and around 16% for the absolute humidifier at 0 mg/L added humidity. Simulations also suggest that bronchial obstruction increases the deposited dose, whereas alveolar enlargement decreases the deposited dose. Conclusions: The delivered dose and hence the simulated lung deposited dose was improved by allowing independent humidity control during nebulization for adult invasive mechanical ventilation.
      pubtype: Academic Journal
      doctype:
        CEU
        pictorial
        research
        tables/charts
        Journal Article
      ougenre: Article
    language: English
    refInfo:
    holdings:
      @attributes:
        islocal: N