Uso del sistema imagen portal electrónica iViewGT para control de calidad del haz de radiación de un acelerador lineal de uso clínico.
The reference state of the iViewGT aSi EPID attached to our Elekta precise linac was established following the recommendation of AAPM. The panel was tested for both available energies: 6 MV and 15 MV. The resolution of the panel is evaluated as the Fourier transform of the edge spread function (ESF)...
| Published in: | Revista Cubana de Física Vol. 27; no. 2B; pp. 251 - 257 |
|---|---|
| Main Authors: | , |
| Format: | Article |
| Published: |
Universidad de La Habana
2010
|
| Subjects: | |
| Online Access: | View this record in EBSCOhost |
| fields | @attributes: recordID: 1 pdfLink: plink: https://search.ebscohost.com/login.aspx?direct=true&db=lth&AN=74457253&site=ehost-live header: @attributes: shortDbName: lth uiTerm: 74457253 longDbName: MedicLatina uiTag: AN controlInfo: bkinfo: jinfo: jid: 02539268 UEW jtl: Revista Cubana de Física issn: 02539268 maglogo: N pubinfo: dt: 2010 vid: 27 iid: 2B pid: 21208 pub: Universidad de La Habana artinfo: ui: 74457253 ppf: 251 ppct: 6 formats: fmt: @attributes: type: P size: 230KB tig: atl: Uso del sistema imagen portal electrónica iViewGT para control de calidad del haz de radiación de un acelerador lineal de uso clínico. aug: au: Silvestre Patallo, I. García Yip, F. affil: Departamento de Radioterapia, Instituto Nacional de Oncología y Radiobiología, La Habana su: Image processing software Linear accelerators in medicine Quality assurance Medical radiology instruments Radiation therapy equipment Evaluation sug: subj: Image processing software Linear accelerators in medicine Quality assurance Medical radiology instruments Radiation therapy equipment Evaluation keyword: Image processing in medical imaging 87.57.C- Quality assurance for radiation therapy equipment: 87.56.Fc Radiation therapy equipment 87.56.-v ab: The reference state of the iViewGT aSi EPID attached to our Elekta precise linac was established following the recommendation of AAPM. The panel was tested for both available energies: 6 MV and 15 MV. The resolution of the panel is evaluated as the Fourier transform of the edge spread function (ESF) of a 1,5 cm thick Pb block placed at the isocenter. Other simple artefacts placed in the beam are useful to determine the geometric precision and linearity, as well as contrast and signal-to-noise ratio. The results are not far from those stated by the supplier and measured by others for this type of detector. It was also established a QC protocol to verify the functionality of the Elekta acquisition software (iViewGT). The described procedures serve as guides for the acceptance and commissioning and the indicators, are used as reference for periodic QC of the EPID. The feasibility of using iViewGT for beam quality control was demonstrated. From an image (16 bits and 1024x1024 pixels) acquired by a short exposure we obtained symmetry, flatness, penumbra and field size. The results were validated against the system "Hermes", used in the weekly quality control. The constancy of the beam quality is evaluated by placing a 5mm thick Pb block on the panel surface. An exposure acquired with metallic pointers aiming at the centre of the crosswire is useful to determine the coincidence of light/ radiation beam centre. We developed procedures to analyse and process the images using a combination of packages available in the web: ImageJ and DoseLab. Se estableció el estado de referencia y el Programa de Garantía de Calidad (PGC) del Sistema de Imagen Portal Electrónica (EPID) iViewGT, compuesto de un panel plano de Si-amorfo, para la evaluación de su desempeño en el tiempo. Se demuestra la factibilidad de su uso como detector para la realización de las pruebas del PGC de los haces de fotones de 6 y 15 MV de un acelerador lineal Elekta Precise de uso clínico. Se describe el procedimiento para evaluar la resolución del panel a través de la MTF como la transformada de Fourier de la Función de Borde (ESF) de un bloque de plomo. Interponiendo en el haz de fotones otros artefactos simples, se determinan la precisión y linealidad geométricas, así como el contraste y la razón señal-ruido. Se diseña un procedimiento para verificar la funcionalidad del software de adquisición (iView) de Elekta. Se demostró la utilidad del iViewGT con fines de controles de calidad del haz de radiación. De la imagen (16 bits y 1024x1024 píxeles) adquirida mediante una corta exposición se determinan la planitud, simetría, penumbra y tamaño del campo. Los resultados fueron validados contra el sistema Hermes (diodos) utilizado en el control semanal. La constancia de la calidad del haz se evalúa colocando un bloque de Pb de 5mm de espesor sobre el panel. Una exposición adquirida con marcadores metálicos que apuntan al centro del retículo sirve para determinar la coincidencia del centro del haz de luz con el del haz de radiación. Para el análisis y procesamiento de las imágenes se crearon procedimientos utilizando una combinación de los paquetes disponibles en la web: ImageJ y DoseLab. pubtype: Academic Journal doctype: Article src: R language: Spanish refInfo: copyright: @attributes: flag: Y custom: Copyright of Revista Cubana de Física is the property of Universidad de La Habana and its content may not be copied or emailed to multiple sites without the copyright holder's express written permission. Additionally, content may not be used with any artificial intelligence tools or machine learning technologies. However, users may print, download, or email articles for individual use. item: Revista Cubana de Física holder: Universidad de La Habana dt: @attributes: year: 2010 holdings: @attributes: islocal: N |
|---|