The use of hyfrecation in the treatment of Verrucae, a case study.

An old but good definition of hyfrecation is; a form of high frequency eiectrosurgery where a low powered electrical impulses are delivered to the surface of the skin via a blunt or sharp electrode. (Sebben 1988). The application of the electrode has differing effect on the skin, either electrodessi...

Full description

Bibliographic Details
Published in:Podiatry Review Vol. 69; no. 2; pp. 4 - 6
Main Authors: Sullivan, Robert, O'Flynn, Deirdre
Format: case study pictorial Journal Article
Published: Institute of Podiatrists Mar/Apr2012
Online Access:View this record in EBSCOhost
fields @attributes:
  recordID: 1
pdfLink:
plink: https://search.ebscohost.com/login.aspx?direct=true&db=ccm&AN=74557307&site=ehost-live
header:
  @attributes:
    shortDbName: ccm
    uiTerm: 74557307
    longDbName: CINAHL Complete
    uiTag: AN
  controlInfo:
    bkinfo:
    dissinfo:
    jinfo:
      jid:
        17563291
        62EP
      jtl: Podiatry Review
      issn: 17563291
      maglogo: N
    pubinfo:
      dt: Mar/Apr2012
      vid: 69
      iid: 2
      pid: 30273
      pub: Institute of Podiatrists
    artinfo:
      ui:
        74557307
        74557307
        108179769
        74557307
      ppf: 4
      ppct: 2
      formats:
        fmt:
          @attributes:
            type: P
      tig:
        atl: The use of hyfrecation in the treatment of Verrucae, a case study.
      aug:
        au:
          Sullivan, Robert
          O'Flynn, Deirdre
      sug:
        subj:
          Electrosurgery Methods
          Warts Surgery
          Adult
          Male
          Nerve Block
          Tibial Nerve
          Treatment Outcomes
          Adult: 19-44 years
          Male
      ab: An old but good definition of hyfrecation is; a form of high frequency eiectrosurgery where a low powered electrical impulses are delivered to the surface of the skin via a blunt or sharp electrode. (Sebben 1988). The application of the electrode has differing effect on the skin, either electrodessication, where the tissue is coagulated and desiccated without burning or electrofulguration where the skin is burnt causing carbonisation of the tissue. Current literature shows that hyfrecation itself does not interfere with nerve conduction due to its high frequency (Plant 2002, Gupta 2003, Fernando et al 2004), pain is caused by the heat generated in the tissue. When hyfrecation is used on Verrucae the area must be well anesthetised using a suitable local anaesthetic and appropriate infiltration. It is the author's experience that the lesion is best pared down prior to the use of the hyfrecator. Both techniques of electrodessication and electrofulguration are used to gain adequate access and depth in the removal of the Verrucae. Hyfrecation is contraindicated in patients with a pacemaker as there is the potential for interference for the frequency of the pulse generated (Bridenstine 1998).