The Accuracy of Matching Target Insertion Gains With Open-Fit Hearing Aids.

Purpose: To assess the accuracy with which target insertion gains were matched for a single type of open-fit hearing aid, both on initial fitting and after adjustment. Method: The hearing aids were fitted using the first-fit setting in the programming software and the target formula was selected as...

Descripción completa

Detalles Bibliográficos
Publicado en:American Journal of Audiology Vol. 21; no. 2; pp. 175 - 181
Autores principales: Aazh, Hashir, Moore, Brian C. J., Prasher, Deepak
Formato: research tables/charts Journal Article
Publicado: American Speech-Language-Hearing Association Dec2012
Acceso en línea:Ver este registro en EBSCOhost
fields @attributes:
  recordID: 1
pdfLink:
plink: https://search.ebscohost.com/login.aspx?direct=true&db=ccm&AN=84510838&site=ehost-live
header:
  @attributes:
    shortDbName: ccm
    uiTerm: 84510838
    longDbName: CINAHL Complete
    uiTag: AN
  controlInfo:
    bkinfo:
    dissinfo:
    jinfo:
      jid:
        10590889
        5KS
      jtl: American Journal of Audiology
      issn: 10590889
      maglogo: N
    pubinfo:
      dt: Dec2012
      vid: 21
      iid: 2
      pid: 42
      pub: American Speech-Language-Hearing Association
      place: Rockville, Maryland
    artinfo:
      ui:
        84510838
        104404239
        104404239
        10.1044/1059-0889(2012/11-0008)
        84510838
      ppf: 175
      ppct: 6
      formats:
        fmt:
          @attributes:
            type: P
      tig:
        atl: The Accuracy of Matching Target Insertion Gains With Open-Fit Hearing Aids.
      aug:
        au:
          Aazh, Hashir
          Moore, Brian C. J.
          Prasher, Deepak
        affil: Royal Surrey County Hospital National Health Service Foundation Trust, United Kingdom
      sug:
        subj:
          Hearing Aids
          Hearing Aid Fitting
          Equipment Design
          Human
          Male
          Female
          Adult
          Middle Age
          Aged
          Aged, 80 and Over
          Descriptive Statistics
          Effect Size
          Adult: 19-44 years
          Middle Aged: 45-64 years
          Aged: 65+ years
          Aged, 80 & over
          Male
          Female
      ab: Purpose: To assess the accuracy with which target insertion gains were matched for a single type of open-fit hearing aid, both on initial fitting and after adjustment. Method: The hearing aids were fitted using the first-fit setting in the programming software and the target formula was selected as NAL-NL1. The difference between the real ear insertion gain (REIG) and the NAL-NL1 target REIG was recorded. The initial fitting was considered acceptable if the difference was less than 10 dB at all frequencies. If an initial fitting was not acceptable, the frequency-gain response was modified. The difference between the final REIG and the NAL-NL1 target REIG was recorded as final target mismatch. Results: Of the 51 initial fittings, 36 (71%) failed to achieve a match within ±10 dB of the NAL-NL1 insertion gain target at 1 or more frequencies between 0.25 and 4 kHz. After the authors adjusted the frequency-gain response of the hearing aids, only 9 fittings (18%) failed to achieve a match. Conclusion: These outcomes suggest that target insertion gains for the open-fit hearing aids used here are rarely achieved with a first fitting but can usually be achieved through adjustments based on REIG measurements.
      pubtype: Academic Journal
      doctype:
        research
        tables/charts
        Journal Article
      ougenre: Article
    language: English
    refInfo:
    holdings:
      @attributes:
        islocal: N