Impact of implementing smart infusion pumps in a pediatric intensive care unit.

Purpose. The impact of smart infusion pumps on the interception of errors in the programming of i.v. drug administrations on a pediatric intensive care unit (PICU) is investigated. Methods. A prospective observational intervention study was conducted in the PICU of a hospital in Madrid, Spain, to es...

Full description

Bibliographic Details
Published in:American Journal of Health-System Pharmacy Vol. 70; no. 21; pp. 1897 - 1907
Main Authors: Manrique-Rodríguez, Silvia, Sánchez-Galindo, Amelia C., López-Herce, Jesús, Calleja-Hernández, Miguel Ãngel, Martínez-Martínez, Fernando, Iglesias-Peinado, Irene, Carrillo-Ãlvarez, Ãngel, Sanjurjo Sáez, María, Fernández-Llamazares, Cecilia M.
Format: research tables/charts Journal Article
Published: Oxford University Press / USA 11/1/2013
Online Access:View this record in EBSCOhost
Description
Summary:Purpose. The impact of smart infusion pumps on the interception of errors in the programming of i.v. drug administrations on a pediatric intensive care unit (PICU) is investigated. Methods. A prospective observational intervention study was conducted in the PICU of a hospital in Madrid, Spain, to estimate the patient safety benefits resulting from the implementation of smart pump technology (Alaris System, CareFusion, San Diego, CA). A systematic analysis of data stored by the devices during the designated study period (January 2010-June 2011) was conducted using the system software (Guardrails CQI Event Reporter, CareFusion). The severity of intercepted errors was independently classified by a group of four clinical pharmacists and a group of four intensive care pediatricians; analyses of intragroup and intergroup agreement in perceptions of severity were performed. Results. During the 17-month study period, the overall rate of user compliance with the safety software was 78%. The use of smart pump technology resulted in the interception of 92 programming errors, 84% of which involved analgesics, antiinfectives, inotropes, and sedatives. About 97% of the errors resulted from user programming of doses or infusion rates above the hard limits defined in the smart pump drug library. The potential consequences of the intercepted errors were considered to be of moderate, serious, or catastrophic severity in 49% of cases. Conclusion. The use of smart pumps in a PICU improved patient safety by enabling the interception of infusion programming errors that posed the potential for severe injury to pediatric patients.