Coronary CT angiography: image quality, diagnostic accuracy, and potential for radiation dose reduction using a novel iterative image reconstruction technique-comparison with traditional filtered back projection.

Objectives: To compare image noise, image quality and diagnostic accuracy of coronary CT angiography (cCTA) using a novel iterative reconstruction algorithm versus traditional filtered back projection (FBP) and to estimate the potential for radiation dose savings.Methods: Sixty five consecutive pati...

Descripción completa

Detalles Bibliográficos
Publicado en:European Radiology Vol. 21; no. 10; pp. 2130 - 2139
Autores principales: Moscariello A, Takx RA, Schoepf UJ, Renker M, Zwerner PL, O'Brien TX, Allmendinger T, Vogt S, Schmidt B, Savino G, Fink C, Bonomo L, Henzler T, Moscariello, Antonio, Takx, Richard A P, Schoepf, U Joseph, Renker, Matthias, Zwerner, Peter L, O'Brien, Terrence X, Allmendinger, Thomas
Formato: research Journal Article
Publicado: Springer Nature Oct2011
Acceso en línea:Ver este registro en EBSCOhost
Descripción
Sumario:Objectives: To compare image noise, image quality and diagnostic accuracy of coronary CT angiography (cCTA) using a novel iterative reconstruction algorithm versus traditional filtered back projection (FBP) and to estimate the potential for radiation dose savings.Methods: Sixty five consecutive patients (48 men; 59.3 ± 7.7 years) prospectively underwent cCTA and coronary catheter angiography (CCA). Full radiation dose data, using all projections, were reconstructed with FBP. To simulate image acquisition at half the radiation dose, 50% of the projections were discarded from the raw data. The resulting half-dose data were reconstructed with sinogram-affirmed iterative reconstruction (SAFIRE). Full-dose FBP and half-dose iterative reconstructions were compared with regard to image noise and image quality, and their respective accuracy for stenosis detection was compared against CCA.Results: Compared with full-dose FBP, half-dose iterative reconstructions showed significantly (p = 0.001 - p = 0.025) lower image noise and slightly higher image quality. Iterative reconstruction improved the accuracy of stenosis detection compared with FBP (per-patient: accuracy 96.9% vs. 93.8%, sensitivity 100% vs. 100%, specificity 94.6% vs. 89.2%, NPV 100% vs. 100%, PPV 93.3% vs. 87.5%).Conclusions: Iterative reconstruction significantly reduces image noise without loss of diagnostic information and holds the potential for substantial radiation dose reduction from cCTA.