Automatic implantable cardioverter/defibrillator: inadvertent discharges during permanent pacemaker magnet tests.

A patient with an automatic impIantable cardioverter defibrillator (AICD)(R) received two inadvertent shacks when a magnet was placed over the pacer during a routine permanent pacer check. Analysis of the rhythm strip suggested that both patients' QRS complexes (133 beats/minute) and asynchronous pa...

Descripción completa

Detalles Bibliográficos
Publicado en:Pacing & Clinical Electrophysiology Vol. 10; no. 3; pp. 579 - 583
Autores principales: Kim SG, Furman S, Matos JA, Waspe LE, Brodman R, Fisher JD
Formato: Journal Article
Publicado: Wiley-Blackwell May1987
Acceso en línea:Ver este registro en EBSCOhost
Descripción
Sumario:A patient with an automatic impIantable cardioverter defibrillator (AICD)(R) received two inadvertent shacks when a magnet was placed over the pacer during a routine permanent pacer check. Analysis of the rhythm strip suggested that both patients' QRS complexes (133 beats/minute) and asynchronous pacer artifacts (70 beats/minute) were counted by the AICD sensing system and exceeded the rate criteria of 153 beats/minute. This resulted in shocks from the AICD during sinus rhythm at 133 beats/minute. To avoid possible inadvertent shocks, an AICD should be deactivated while a magnet is placed over the pacemaker during a permanent pacer check.