Automatic implantable cardioverter/defibrillator: inadvertent discharges during permanent pacemaker magnet tests.

A patient with an automatic impIantable cardioverter defibrillator (AICD)(R) received two inadvertent shacks when a magnet was placed over the pacer during a routine permanent pacer check. Analysis of the rhythm strip suggested that both patients' QRS complexes (133 beats/minute) and asynchronous pa...

Full description

Bibliographic Details
Published in:Pacing & Clinical Electrophysiology Vol. 10; no. 3; pp. 579 - 583
Main Authors: Kim SG, Furman S, Matos JA, Waspe LE, Brodman R, Fisher JD
Format: Journal Article
Published: Wiley-Blackwell May1987
Online Access:View this record in EBSCOhost
fields @attributes:
  recordID: 1
pdfLink:
plink: https://search.ebscohost.com/login.aspx?direct=true&db=ccm&AN=106034530&site=ehost-live
header:
  @attributes:
    shortDbName: ccm
    uiTerm: 106034530
    longDbName: CINAHL Complete
    uiTag: AN
  controlInfo:
    bkinfo:
    dissinfo:
    jinfo:
      jid:
        01478389
        4F8
      jtl: Pacing & Clinical Electrophysiology
      issn: 01478389
      maglogo: Y
    pubinfo:
      dt: May1987
      vid: 10
      iid: 3
      pid: 480
      pub: Wiley-Blackwell
      place: Malden, Massachusetts
    artinfo:
      ui:
        106034530
        2009407478
        NLM2440009
        106034530
      ppf: 579
      ppct: 4
      formats:
        fmt:
          @attributes:
            type: P
      tig:
        atl: Automatic implantable cardioverter/defibrillator: inadvertent discharges during permanent pacemaker magnet tests.
      aug:
        au:
          Kim SG
          Furman S
          Matos JA
          Waspe LE
          Brodman R
          Fisher JD
      sug:
      ab: A patient with an automatic impIantable cardioverter defibrillator (AICD)(R) received two inadvertent shacks when a magnet was placed over the pacer during a routine permanent pacer check. Analysis of the rhythm strip suggested that both patients' QRS complexes (133 beats/minute) and asynchronous pacer artifacts (70 beats/minute) were counted by the AICD sensing system and exceeded the rate criteria of 153 beats/minute. This resulted in shocks from the AICD during sinus rhythm at 133 beats/minute. To avoid possible inadvertent shocks, an AICD should be deactivated while a magnet is placed over the pacemaker during a permanent pacer check.
      pubtype: Academic Journal
      doctype: Journal Article
      ougenre: Article
    language: English
    refInfo:
    holdings:
      @attributes:
        islocal: N