The automatic implantable cardioverter/ defibrillator: transesophageal atrial pacing discloses the potential for erroneous discharges.

We report the occurrence of erroneous discharge from an implanted automatic cardioverter/defibrillator during transesophageal atrial pacing. Transesophageal pacing was performed as part of a study protocol on the inducibility of ventricular tachycardia from the atrium in patients with ischemic heart...

Descripción completa

Detalles Bibliográficos
Publicado en:Pacing & Clinical Electrophysiology Vol. 11; no. 4; pp. 419 - 423
Autores principales: Critelli G, Monda V, Scherillo M, Coltorti F, Greco C, Reale A
Formato: case study Journal Article
Publicado: Wiley-Blackwell Apr1988
Acceso en línea:Ver este registro en EBSCOhost
Descripción
Sumario:We report the occurrence of erroneous discharge from an implanted automatic cardioverter/defibrillator during transesophageal atrial pacing. Transesophageal pacing was performed as part of a study protocol on the inducibility of ventricular tachycardia from the atrium in patients with ischemic heart disease. At on induced heart rate of 166 beats per minute (a value just above the cut-off rate of the device), the cardioverter/defibrillator was triggered. This observation suggests that transesophageal atrial pacing could be utilized to disclose the potential [or spurious discharges in the event of fast atrial rhythms in patients with the automatic implantable cardioverter/defibrillator.