Delayed detection of ventricular tachycardia in a dual chamber rate adaptive pacing implantable cardioverter defibrillator: a case of intradevice interaction.

Ventricular tachycardia detection is delayed due to concomitant programming of rate adaptive dual chamber pacing. Underlying timing cycle constraints as well as programming precautions and pitfalls are reviewed.

Detalles Bibliográficos
Publicado en:Pacing & Clinical Electrophysiology Vol. 27; no. 8; pp. 1164 - 1167
Autor principal: Shalaby AA
Formato: case study tables/charts Journal Article
Publicado: Wiley-Blackwell Aug2004
Acceso en línea:Ver este registro en EBSCOhost
fields @attributes:
  recordID: 1
pdfLink:
plink: https://search.ebscohost.com/login.aspx?direct=true&db=ccm&AN=106526697&site=ehost-live
header:
  @attributes:
    shortDbName: ccm
    uiTerm: 106526697
    longDbName: CINAHL Complete
    uiTag: AN
  controlInfo:
    bkinfo:
    dissinfo:
    jinfo:
      jid:
        01478389
        4F8
      jtl: Pacing & Clinical Electrophysiology
      issn: 01478389
      maglogo: Y
    pubinfo:
      dt: Aug2004
      vid: 27
      iid: 8
      pid: 480
      pub: Wiley-Blackwell
      place: Malden, Massachusetts
    artinfo:
      ui:
        106526697
        106526697
        2009036090
        10.1111/j.1540-8159.2004.00603.x
        NLM15305971
        106526697
      ppf: 1164
      ppct: 3
      formats:
        fmt:
          @attributes:
            type: P
      tig:
        atl: Delayed detection of ventricular tachycardia in a dual chamber rate adaptive pacing implantable cardioverter defibrillator: a case of intradevice interaction.
      aug:
        au: Shalaby AA
        affil: Assistant Professor of Medicine, University of Pittsburgh, Director of Cardiac Electrophysiology, Division of Cardiology, Pittsburgh VA Healthcare System 111©) University Drive C, Pittsburgh PA 15240; Alaa.Shalaby@med.va.gov
      sug:
        subj:
          Cardiac Pacing, Artificial
          Defibrillators, Implantable
          Equipment Failure
          Tachycardia, Ventricular Diagnosis
          Male
          Middle Age
          Time Factors
          Middle Aged: 45-64 years
          Male
      ab: Ventricular tachycardia detection is delayed due to concomitant programming of rate adaptive dual chamber pacing. Underlying timing cycle constraints as well as programming precautions and pitfalls are reviewed.
      pubtype: Academic Journal
      doctype:
        case study
        tables/charts
        Journal Article
      ougenre: Article
    language: English
    refInfo:
    holdings:
      @attributes:
        islocal: N