Delayed detection of ventricular tachycardia in a dual chamber rate adaptive pacing implantable cardioverter defibrillator: a case of intradevice interaction.

Ventricular tachycardia detection is delayed due to concomitant programming of rate adaptive dual chamber pacing. Underlying timing cycle constraints as well as programming precautions and pitfalls are reviewed.

Detalles Bibliográficos
Publicado en:Pacing & Clinical Electrophysiology Vol. 27; no. 8; pp. 1164 - 1167
Autor principal: Shalaby AA
Formato: case study tables/charts Journal Article
Publicado: Wiley-Blackwell Aug2004
Acceso en línea:Ver este registro en EBSCOhost