Myocardial multilayer strain does not provide additional value for detection of myocardial viability assessed by SPECT imaging over and beyond standard strain.

Background: The aim of this study was to evaluate the value of multilayer strain analysis to the assessment of myocardial viability (MV) through the comparison of both speckle tracking echocardiography and single‐photon emission computed tomography (SPECT) imaging. We also intended to determine whic...

Descripción completa

Detalles Bibliográficos
Publicado en:Echocardiography Vol. 35; no. 9; pp. 1300 - 1310
Autores principales: Orloff, Elisabeth, Fournier, Pauline, Bouisset, Frédéric, Moine, Thomas, Cournot, Maxime, Elbaz, Meyer, Carrié, Didier, Galinier, Michel, Lairez, Olivier, Cognet, Thomas
Formato: research tables/charts Journal Article
Publicado: Wiley-Blackwell Sep2018
Acceso en línea:Ver este registro en EBSCOhost
Descripción
Sumario:Background: The aim of this study was to evaluate the value of multilayer strain analysis to the assessment of myocardial viability (MV) through the comparison of both speckle tracking echocardiography and single‐photon emission computed tomography (SPECT) imaging. We also intended to determine which segmental longitudinal strain (LS) cutoff value would be optimal to discriminate viable myocardium. Methods: We included 47 patients (average age: 61 ± 11 years) referred to our cardiac imaging center for MV evaluation. All patients underwent transthoracic echocardiography with measures of LS, SPECT, and coronary angiography. Results: In all, 799 segments were analyzed. We correlated myocardial tracer uptake by SPECT with sub‐endocardial, sub‐epicardial, and mid‐segmental LS values with r = .514 P < .0001, r = .501 P < .0001, and r = .520 P < .0001, respectively. The measurements of each layer strain (sub‐endocardial, sub‐epicardial, and mid) had the same performance to predict MV viability as defined by SPECT with areas under curve of 0.819 [0.778–0.861, P < .0001], 0.809 [0.764–0.854, P < .0001], and 0.817 [0.773–0.860, P < .0001], respectively. The receiver‐operating characteristic analysis yielded a cutoff value of −6.5% for mid‐segmental LS with a sensitivity of 76% and specificity of 76% to predict segmental MV as defined by SPECT. Conclusions: Multilayer strain analysis does not evaluate MV with more accuracy than standard segmental LS analysis.