The Effect of Nail Polish on Pulseoximeter's Measurements in Healthy Individuals.

Background: Pulse oximetry is considered as the most valuable non-invasive method for continuous oxygen saturation (SpO2) monitoring in patients with hypoxemia risk. Aim: The aim of this study to determine whether nail polish affected peripheral capillary hemoglobin oxygen saturation measurements in...

Descripción completa

Detalles Bibliográficos
Publicado en:International Journal of Caring Sciences Vol. 12; no. 2; pp. 1144 - 1148
Autores principales: Yikar, Seda Karacay, Arslan, Sevban, Nazik, Evşen
Formato: research tables/charts Journal Article
Publicado: International Journal of Caring Sciences May-Aug2019
Acceso en línea:Ver este registro en EBSCOhost
fields @attributes:
  recordID: 1
pdfLink:
plink: https://search.ebscohost.com/login.aspx?direct=true&db=ccm&AN=138636107&site=ehost-live
header:
  @attributes:
    shortDbName: ccm
    uiTerm: 138636107
    longDbName: CINAHL Complete
    uiTag: AN
  controlInfo:
    bkinfo:
    dissinfo:
    jinfo:
      jid:
        17915201
        B92O
      jtl: International Journal of Caring Sciences
      issn: 17915201
      maglogo: N
    pubinfo:
      dt: May-Aug2019
      vid: 12
      iid: 2
      pid: 64496
      pub: International Journal of Caring Sciences
    artinfo:
      ui:
        138636107
        138636107
        138636107
        138636107
      ppf: 1144
      ppct: 4
      formats:
        fmt:
          @attributes:
            type: P
      tig:
        atl: The Effect of Nail Polish on Pulseoximeter's Measurements in Healthy Individuals.
      aug:
        au:
          Yikar, Seda Karacay
          Arslan, Sevban
          Nazik, Evşen
        affil: Research Assistant, Cukurova University, Faculty of Health Sciences, Obstetrics and Gynecology Nursing Department, Adana, Turkey
      sug:
        subj:
          Oxygen Saturation Evaluation
          Pulse Oximetry
          Cosmetics Adverse Effects
          Nail Care
          Human
          Female
          Adolescence
          Adult
          Quasi-Experimental Studies
          Nursing Role
          Descriptive Statistics
          Portable Equipment
          Data Analysis Software
          T-Tests
          Color
          Adolescent: 13-18 years
          Adult: 19-44 years
          Female
      ab: Background: Pulse oximetry is considered as the most valuable non-invasive method for continuous oxygen saturation (SpO2) monitoring in patients with hypoxemia risk. Aim: The aim of this study to determine whether nail polish affected peripheral capillary hemoglobin oxygen saturation measurements in healthy individuals Methodology: This is a semi-experimental study.. This research was conducted with the 80 female who 18 age and above and not having any health problem, not having any communication problem, non-smoker, agreed to participate in study at a Nursing Department from January 2015 to June 2015 were included. The right hand of the participants was selected as the hand to receive nail polish for the experiment, while the left hand was used as a control without nail polish. A portable pulse oximeter was used for oxygen saturation measurement. Results; There was statistical significant difference between SpO2 measurements performed on the right and left thumbs (color pearl, p=.000). Similarly, statistically significant differences were found between measurements of right and left hand index finger (color red, p=.000), middle finger (color blue, p=.000), ring finger (color brown, p=.000) and small finger (color green, p=.000). Conclusion: As a result, all the nail polish colors caused a reduction in the measurement of SpO2. Nurses may be recommended to remove nail polish of the patients before measuring oxygen saturation on the finger
      pubtype: Academic Journal
      doctype:
        research
        tables/charts
        Journal Article
      ougenre: Article
    language: English
    refInfo:
    holdings:
      @attributes:
        islocal: N