Rapid point of care testing for four bacterial sexually transmitted infections using the portable isothermal loop-mediated nucleic acid amplification eazyplex platform.

Purpose: To analyze sensitivity and specificity of the rapid point-of-care (POC) eazyplex testing platform for bacterial sexually transmitted infections (STI) among men who have sex with men (MSM). Methods: 272 anal, urethral, and pharyngeal swabs collected from 153 MSM were tested by both the eazyp...

Descripción completa

Detalles Bibliográficos
Publicado en:Infection Vol. 51; no. 4; pp. 1141 - 1146
Autores principales: Esitgen Germaner, Ece T., Wassill, Lars, Dichtl, Karl, Roider, Julia, Seybold, Ulrich
Formato: research tables/charts Journal Article
Publicado: Springer Nature Aug2023
Acceso en línea:Ver este registro en EBSCOhost
Descripción
Sumario:Purpose: To analyze sensitivity and specificity of the rapid point-of-care (POC) eazyplex testing platform for bacterial sexually transmitted infections (STI) among men who have sex with men (MSM). Methods: 272 anal, urethral, and pharyngeal swabs collected from 153 MSM were tested by both the eazyplex platform and an in-house PCR or culture in the university microbiology reference laboratory. Results: Compared to the reference diagnostic method, the overall sensitivity/specificity of eazyplex was 88%/98% for N. gonorrhoeae, 82%/100% for C. trachomatis, 70%/ > 99% for U. urealyticum, and 85%/98% for M. hominis, respectively. Sensitivity for N. gonorrhoeae and U. urealyticum in urethral samples was 100%. Conclusion: With good to very good sensitivity depending on the sampling site and pathogen as well as very good specificity overall the eazyplex platform is a useful rapid diagnostic method for POC bacterial STI-testing especially for N. gonorrhoeae and C. trachomatis, allowing for almost immediate treatment initiation.